Fiber Amplifier 

Chapter 5: Operation

 

TTN118382-D02, Rev A, November 18, 2016

 

Page 8 

5.2. 

Turning On the Amplifier 

1. 

Please consult with your organization’s laser safety officer regarding proper operation of the  amplifier at 
your institution.  

LASER WARNING 

The amplifier generates amplified spontaneous emission (ASE) even without any optical input. The 

power level of the ASE can be as high as 30 mW. Therefore, all laser safety measures should be 

followed even if the input light to the amplifier has been disconnected or switched off. Additionally, 

it should be noted that connecting a laser to the output port of the amplifier by mistake can cause 

damage to the laser due to the ASE light getting coupled into the laser source. 

2.  Turn the POWER key switch clockwise. 

The LCD display will scroll “Thorlabs EDFA” across the screen, 

followed by the software revision number. 

3.  Make sure the interlock input is short-circuited. See page 10 for detailed instructions. 

4.  Press  and  release  the  ENABLE  switch  to  activate  the  fiber  amplifier.  There  will  be  an  approximately 

3 second delay before the  fiber amplifier powers up. During this time, the ENABLE indicator will light up 
and blink rapidly. 

5.  Readings for the fiber amplifier pump current (in mA) and temperature (in °C) will be activated when the 

unit turns on. The temperature will typically require 1 to 2 minutes to stabilize.  

5.3. 

Viewing Information 

The EDFA100S and EDFA100P use a single four-quadrant LCD to display and access information. At any time, 
display variables can be adjusted by simply rotating the control knob located to the left of the display. The following 
information will be available: 

 

Top Left 

– Indicates device type (EDFA).   

 

Top Right 

– Indicates the amplification band of the 

device. This is set to 1550 nm for the EDFA100S and 
EDFA100P fiber amplifiers. 

 

Bottom Left 

– Indicates the pump current.   

 

Bottom  Right 

–  Indicates  the  actual  temperature  of  the  pump  laser,  in 

C.  The  system  defaults  to  a 

temperature  that  is  set  at  the  factory.  The  temperature  control  is  always  active  and  may  require  1  to  2 
minutes to fully stabilize. 

5.4. 

Adjusting the Fiber Amplifier Gain 

The fiber amplifier gain and output power are controlled by varying the pump current. The system specifications 
and the test data provided for each unit show the performance at a pump temperature of 25 °C and the maximum 
pump current of 1000 mA. A typical power scaling plot, at two different input power levels of 0 dBm and -20 dBm, 
is shown in Figure 2 below. This plot can be used as a reference to adjust the gain and power of the amplifier by 
varying the pump current. 

To adjust the pump current, press the adjustment knob once. The bottom left indicator on the LCD screen, which 
shows the current in mA, will start blinking when the control knob is pressed. Adjust the control knob until the desired 
current is achieved. Adjusting the knob clockwise will  incrementally increase the pump diode drive current, up to 
the maximum drive current of 1000 mA. Adjusting the knob counterclockwise will incrementally decrease the drive 
current. On system shutdown, the current setting will be remembered.  

Note:

 The adjustment knob utilizes an intelligent speed control. Adjusting the knob slowly will increment values in 

small steps, while adjusting the knob quickly will cause larger movements. This allows both fine and coarse control.

 

 

Содержание EDFA100P

Страница 1: ...EDFA100S and EDFA100P Erbium Doped Fiber Amplifiers Operating Manual ...

Страница 2: ...ber Amplifier Gain 8 5 5 Turning the Fiber Amplifier Off 9 Making Safety Interlock Connections 10 Remote Communications 11 7 1 Installing the USB Drivers 11 7 2 Command Line Interface 11 7 3 Keywords Commands and Queries 12 Troubleshooting 13 General Maintenance 14 9 1 Cleaning 14 9 2 Connector Cleaning 14 Specifications 16 Mechanical Drawings 18 Regulatory 19 12 1 Waste Treatment is Your Own Resp...

Страница 3: ...ual or on your device Symbol Description Direct Current Alternating Current Both Direct and Alternating Current Earth Ground Terminal Protective Conductor Terminal Frame or Chassis Terminal Equipotentiality On Supply Off Supply In Position of a Bi Stable Push Control Out Position of a Bi Stable Push Control Caution Risk of Electric Shock Caution Hot Surface Caution Risk of Danger Warning Visible o...

Страница 4: ...he amplifier Laser radiation is emitted from the output port when a light source is connected to the input port even if the amplifier is powered off Use of controls or adjustments or performance of procedures other than those specified herein may result in hazardous radiation exposure The unit is supplied with a region specific power cord If using your own power cord make sure it is IEC 320 compat...

Страница 5: ...For added safety there is an interlock connector located on the rear panel that must be shorted in order for the output to be enabled This can easily be configured to be triggered by doors to disable the fiber amplifier in unsafe conditions The power switch is a key lock system to prevent accidental or unwanted use An enable button must be set to activate the amplifier and a green LED indicator di...

Страница 6: ...to the AC receptacle and plug the unit in 4 3 Initial Setup 1 Set the unit on a dry level working surface 2 Make sure the POWER key switch on the front of the unit is in the OFF position key perpendicular to working surface 3 Plug the female end of the provided AC line cord into the IEC input receptacle on the rear of the unit Plug the male end into a properly grounded AC socket 4 Install the inte...

Страница 7: ... Switch Key can only be removed when the amplifier is off Control Knob Switch Adjusts pump laser current and temperature Remote Interlock Input See page 10 for connection logic USB Input Connector Allows full operation through a PC interface see Chapter 7 Fuse Tray See Section 4 2 for replacement details Digital Display with Backlight Displays pump laser temperature and drive current Amplifier Out...

Страница 8: ...cated to the left of the display The following information will be available Top Left Indicates device type EDFA Top Right Indicates the amplification band of the device This is set to 1550 nm for the EDFA100S and EDFA100P fiber amplifiers Bottom Left Indicates the pump current Bottom Right Indicates the actual temperature of the pump laser in C The system defaults to a temperature that is set at ...

Страница 9: ...so be achieved by allowing the display to time out at any point in the process The display will adjust in real time to the new current setting 5 5 Turning the Fiber Amplifier Off Standby Mode By adjusting the control knob fully counterclockwise the pump current will adjust down to the threshold current and then to standby mode Disable Enable Mode The amplifier output can be turned off by pressing ...

Страница 10: ... will not re enable until the ENABLE switch is pressed or until the Set Enable command is entered in the command line interface see Section 7 3 All units shipped from Thorlabs are configured with a shorting device installed in the interlock connector If you are not going to use this feature then you can leave the shorting device installed and the unit will operate normally as described throughout ...

Страница 11: ...nd the power turned on configure the terminal emulator as follows Baud Rate 115 2K Bits Per Second Data Bits 8 Parity None Stop Bits 1 Flow Control None If the connection is correct you will see the following after pressing the Enter key Command error CMD_NOT_DEFINED Followed immediately by the prompt The basic structure of the interface is a keyword followed by either an equals sign or a question...

Страница 12: ... the temperature and current when the arrow keys are pressed Get Step step Returns the increment used to adjust the temperature and current when the arrow keys are pressed Save Parameters save Saves target current and target temperature to EEPROM Values are restored on device startup Get Status statword Returns a string representation of an 8 bit number indicating the device s status The right mos...

Страница 13: ...s please return the unit to Thorlabs for evaluation Unit does not enable when pressing the ENABLE button 1 Make sure the AC line cord is properly plugged in and the key switch is turned to the ON position 2 Check to make sure the interlock jumper is installed on the rear panel See page 10 for details Unit is enabled but there is no output 1 Check to make sure you are using the correct type of fibe...

Страница 14: ...g can be cleaned using a soft slightly damp cloth Avoid using any solvents on or near the unit Keep the vent holes located on the bottom of the unit and on the rear panel free of dust buildup Restricted airflow will cause the temperature controls to operate inefficiently and in extreme cases loss of temperature control 9 2 Connector Cleaning Always clean the ferrule end of your fiber patch cables ...

Страница 15: ...r firmly swipe the connector tip across the exposed cleaning strip The connector tip should be flush against the card surface i e held at a slight angle for FC APC connectors If you need additional connector cleaners Thorlabs offers the FCC 7020 Fiber Cleaning Cloth Spool To use the FBC1 Bulkhead and Connector Cleaner please refer to the instructions shipped with the FBC1 ...

Страница 16: ...r Aligned to Slow Axis Polarization Extinction Ratio N A 25 dB Polarization Dependent Gain 0 2 dB N A Return Loss at Input Port 50 dB Input Output Isolation 30 dB Input Output Fiber Type SMF 28 J9 PM1550 XP Input Output Fiber Connectors FC APC Compatible 2 0 mm Narrow Key a The wavelength range over which the output power at 3 dBm input power does not fall below 18 dBm b Specified at 1550 nm Pleas...

Страница 17: ... mm x 77 7 mm Weight 2 08 kg 4 58 lbs Connections and Controls Interface Control Optical Encoder with Push Button Enable Select Keypad Switch Enable with LED Indicator Power On Key Switch Fiber Connectors FC APC Compatible 2 0 mm Narrow Key Display LCD 16x2 Alphanumeric Characters Input Power Connector IEC Connector Interlock 2 5 mm Mono Jack See Chapter 6 Communications Communications Port USB 2 ...

Страница 18: ...Fiber Amplifier Chapter 11 Mechanical Drawings TTN118382 D02 Rev A November 18 2016 Page 18 Mechanical Drawings Figure 5 EDFA100S Mechanical Drawing Figure 6 EDFA100P Mechanical Drawing ...

Страница 19: ...rlabs products such as Pure OEM products that means assemblies to be built into a unit by the user e g OEM laser driver cards Components Mechanics and optics Left over parts of units disassembled by the user PCB s housings etc If you wish to return a Thorlabs unit for waste recovery please contact Thorlabs or your nearest dealer for further information 12 1 Waste Treatment is Your Own Responsibili...

Страница 20: ...Fiber Amplifier Chapter 13 Declaration of Conformity TTN118382 D02 Rev A November 18 2016 Page 20 Declaration of Conformity ...

Страница 21: ...ope thorlabs com Scandinavia Thorlabs Sweden AB Bergfotsgatan 7 431 35 Mölndal Sweden Tel 46 31 733 30 00 Fax 46 31 703 40 45 www thorlabs com Email scandinavia thorlabs com France Thorlabs SAS 109 rue des Côtes 78600 Maisons Laffitte France Tel 33 0 970 444 844 Fax 33 0 825 744 800 www thorlabs com Email sales fr thorlabs com Brazil Thorlabs Vendas de Fotônicos Ltda Rua Riachuelo 171 São Carlos S...

Страница 22: ...www thorlabs com ...

Отзывы: