© 2020 Thorlabs GmbH
4 Write Your Own Application
45
OUTPut subsystem commands
Command
Description
SCPI
OUTPut[1]
Path to OUTPut subsystem
Vol.2 §15
[:STATe] {ON|1|OFF|0}
Enables (ON) or disables (OFF) LED output
Vol.2 §15.12
[:STATe]?
Returns output state
Vol.2 §15.12
:TERMinal {1|2}
Select output terminal: 1 = 10A/12Pin connector
(LED1); 2 = 2A/4Pin connector (LED2);
:TERMinal?
Returns selected output terminal
:TERMinal:ABORt
Abort any running LED head presence test
:TERMinal{[1]|2}
:TEST
Testing presence and properties of LEDs
[:INITiate]
Initiate presence test procedure on the specified ter-
minal
:STATus?
Query status of the presence test procedure as an
<NF1> value:
0 = The presence test procedure is running
1 = The latest presence test detected no LED
2 = The latest presence test detected a custom LED
without onboard head info memory
3 = The latest presence test detected a LED head
with onboard head info memory
:PROTection
:INTLock[:TRIPped]?
Returns interlock circuit protection tripped (1) or un-
tripped (0)
:TEMPerature
[:DRIVer][:TRIPped]?
Returns driver (console) over temperature protection
tripped tripped (1) or untripped (0)
:HEAD[:TRIPped]?
Returns head over temperature protection tripped
tripped (1) or untripped (0)
SENSe3 LED current sensing subsystem commands
Command
Description
SCPI
SENSe3
Path to LED current sensing
[:CURRent][:DC]
[:DATA]? [{MIN|MAX}]
Returns the measured LED current. Parameters MIN
and MAX return the measurement value range pos-
sible in the currently active driver configuration (de-
pends on the compliance voltage required by the
LED head)
SENSe4 LED voltage sensing subsystem commands
Command
Description
SCPI
SENSe4
Path to LED voltage sensing
[:VOLTage][:DC]
[:DATA]? [{MIN|MAX}]
Returns the measured LED voltage. Parameters MIN
and MAX return the measurement value range pos-
sible.
SENSe5 LED temperature sensing subsystem commands
Command
Description
SCPI
SENSe5
Path to LED temperature sensing
[:TEMPerature]? [{MIN|MAX}]
Returns the measured LED temperature. Parameters
MIN and MAX return the measurement value range
possible.
Содержание DC2200
Страница 1: ...High Power LED Controller DC2200 Operation Manual 2020 ...
Страница 2: ...Copyright 2020 Thorlabs GmbH Version Date 1 6 23 Sep 2020 ...
Страница 26: ... 2020 Thorlabs GmbH 24 DC2200 Click I accept if you do so then Next to continue Click Next to continue ...
Страница 31: ... 2020 Thorlabs GmbH 3 Operating Instruction 29 Click Next and then Install to continue ...
Страница 70: ... 2020 Thorlabs GmbH 68 DC2200 6 2 Dimensions ...
Страница 74: ... 2020 Thorlabs GmbH 72 DC2200 6 7 Certifications and Compliances ...
Страница 78: ...www thorlabs com ...