Chapter 5: Alignment
Page 18
TTN118795-D02
18.
Use the 2 mm hex key to adjust the XY Mount inside the microscope (the SM Patch cable is connected to
the XY Mount) in the X and Y directions to center the maximum beam intensity at the center of the image.
Note:
Adjust the histogram in the ThorImageLS software to visualize the intensity of the beam. Click the
histogram icon
in the
Image Control
panel to display the histogram of the image. To adjust the histogram, click the
Auto
button.
19.
Use the 2 mm hex key to adjust the pinhole position in the X and Y directions to get maximum beam
intensity (a XY Mount is attached to the pinhole inside the microscope).
The fluorescence slide alignment procedure is now complete.