B
B-16
HeartMate III Left Ventricular Assist System Instructions for Use
Field strengths from fixed RF transmitters, such as base stations for radios (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV
broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment (home, office, etc.) due to fixed RF transmitters, an
electromagnetic (EMC) “site survey” (measurement) should be considered. If the measured field
strength (in Volts per meter, V/m) in the location(s) in which the HeartMate III Left Ventricular Assist
System is used exceeds the applicable RF compliance level for the frequency bands above, the
HeatMate III LVAS should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as re-orienting or relocating the
HeartMate III LVAS.
These guidelines may not apply in all situations.
WARNING !
• Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 20 cm (8 inches) to any part of the
HeartMate III LVAS (powered by the Mobile Power Unit), including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
• To help ensure good electromagnetic performance throughout the usable life of the
HeartMate III LVAS, be certain to follow the maintenance schedule and procedures as
described in Equipment Storage and Care.
• Avoid use of the HeartMate III LVAS (and its cables and parts) within 5 cm (2 inches) of a
microwave oven or its cord. Keep all HeartMate III LVAS components and their cables away
from other electronic devices and their cables/cords.
• The HeartMate III Left Ventricular Assist System should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use is necessary, the HeartMate III Left
Ventricular Assist System should be observed to verify normal operation in the configuration
in which it will be used.
• Use of accessories or cables other than those specified or provided by Thoratec Corp. could
result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
• No modification of this equipment is allowed.
Содержание HeartMate III
Страница 1: ...Instructions for Use ...
Страница 2: ......
Страница 3: ...Thoratec Corporation HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM Instructions for Use ...
Страница 4: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ii ...
Страница 6: ...HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE iv ...
Страница 10: ...Contents viii HEARTMATE III LEFT VENTRICULAR ASSIST SYSTEM INSTRUCTIONS FOR USE ...
Страница 12: ...1 Introduction 1 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 28: ...1 Introduction 1 18 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 30: ...2 System Operations 2 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 88: ...3 Powering the System 3 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 112: ...3 Powering the System 3 26 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 166: ...4 System Monitor 4 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 208: ...4 System Monitor 4 44 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 210: ...5 Surgical Procedures 5 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 266: ...6 Patient Care and Management 6 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 330: ...7 Alarms and Troubleshooting 7 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 370: ...8 Equipment Storage and Care 8 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 378: ...A A 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 390: ...B B 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 420: ...C C 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 426: ...C C 8 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 428: ...D D 2 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 448: ...E E 10 HeartMate III Left Ventricular Assist System Instructions for Use ...
Страница 457: ......