OPERATION 
 

WARNING: 1. 

To avoid injury do not use flammable solvents with this 

equipment, as fire or explosion may result. This equipment contains 
components that may ignite such materials. 
 

WARNING: 2.

 Do not introduce oil into this bath as the concentrated 

heat energy may result in Pyrolysis (Chemical Decomposition by 
Heat).

 

 

OPERATING THE HEATER CIRCULATOR 

 

1)  Ensure that the heater circulator is located as previously described, refer to 

the index, LOCATION, Page 2. 

 
2)  Fill the bath with cool *clean water to a minimum of 80mm above the element. 

* - Refer to Page 4 section Maintenance for details on water condition and 
compatibility with stainless steel. 
 

     

Warning: 

This equipment contains components rated 

at 240V AC. Ensure that it is secure and the control 
enclosure will not contact water. 

 
 

3)  Connect the electrical lead into a 10 Amp 240 Volt 50Hz GPO (General 

Purpose Outlet). 

 
4)  Select the desired operating temperature at the control knob  

 
 

Caution:

 Contact with hot fluid/vapours, bath surfaces 

or product under evaluation may result in personal 
injury, ensure appropriate protective equipment and 
procedures are applied at all times. 
 

 

5)  Turn the power outlet (GPO) and the heater circulator switch to the ON 

position and allow sufficient time for the water to stabilise. 

 
6)  When the water has stabilised, check the temperature with a mercury in glass 

thermometer (or similar) and adjust the control knob setting if necessary. 

 
 

7)  Optional – Support Bracket – TUSB 

 
a) 

The bracket is attached at the rear of the heater control case using the 
bolt and washers supplied. 

 

 

b) 

With the bracket attached, position the heater circulator over the edge 
of the bath, with the flat face of the bracket to the inside of the bath. 
Slowly tighten the screwed adjuster until the circular pad is held firm 
against the outside of the bath. 

 

 

Содержание TU-1

Страница 1: ...Thermoline Scientific ABN 80 000 859 129 10 12 Ross Place Wetherill Park N S W 2164 Australia P O Box 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 02 9604 3911 International 61 2 96...

Страница 2: ...ere is a Possibility of personal injury Caution alerts apply when there is a Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT BOXES located thr...

Страница 3: ...t for damage Retain the packaging materials until the equipment has been thoroughly tested Notify the detail of any damage without delay to your supplier or Thermoline Scientific LOCATION Place the eq...

Страница 4: ...controls enclosure Below is suspended the circulating pump housing heating element and sensor The desired operating temperature is set at the control A signal from the sensor is monitored at the contr...

Страница 5: ...mp soft cloth The element and pump housing refer to following note Note Stainless steel is under most conditions extremely resistant to corrosion This is in part due to the addition of chromium and ni...

Страница 6: ...f the power and consult a qualified technician The temperature range is 30 C to 100 C Ambient 5 C Model TU 1 The desired operating temperature is set at the control knob against the scale which is gra...

Страница 7: ...he electrical lead into a 10 Amp 240 Volt 50Hz GPO General Purpose Outlet 4 Select the desired operating temperature at the control knob Caution Contact with hot fluid vapours bath surfaces or product...

Страница 8: ...ng the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to replacing the product or the...

Отзывы: