CAUTION: 

The range of heater 

circulators described here are not to be 
used in conjunction with 

oil

 as Pyrolysis 

(Chemical Decomposition by Heat) will 
occur. 

 

 

 

 

 

WARNING: 

Do not use flammable 

solvents or combustible materials as fire 
or explosion may occur. This equipment  
uses components which may be a 
source of ignition.

 

 

 

FUNCTION 

 
To provide an accurate controlled temperature b30

°

C to +100

°

C.  

Model TU-1: a5° to +100°C.

 

 

PRINCIPLE OF OPERATION 

 
The 

analogue

 range of Heater Circulators are designed to be compact, light and 

efficient heater circulators of water. They are ideally suited to Laboratory Bath 
applications where minimal spacial variation is tolerated. 
 
The analogue controller, and long shaft (240 Volt) electric motor are located within 
the controls enclosure. Below is suspended, the circulating pump housing, heating 
element and sensor. 
 
The desired operating temperature is set at the control. A signal from the sensor is 
monitored at the controller. The resultant output switches power to the heating 
element. 
 
Energy from the element gently heats the water, the amount of energy is determined 
by the sensor. 
 
The circulating pump runs continuously maintaining close temporal uniformity 
throughout the bath. 
 
The heater circulator may be supported on a fabricated bridge (standard with each 
Thermoline Scientific circulating water bath) or from an 

optional 

bath to heater 

bracket/clamp. 
 

Note:

 When working at temperatures above 60°C a suitable bath cover (lid) should 

be used to maintain the operating environment and minimise energy loss. 

 

PERFORMANCE 

 
The operating range for this equipment is +30°C to +100

°

C Model TU-1 is ambient 

+5°C to +100°C, Temporal Stability 0.5°C.

 

 

MAXIMUM TEMPERATURE 

For this product is +100

°

C. 

Содержание TU-1

Страница 1: ...Thermoline Scientific ABN 80 000 859 129 10 12 Ross Place Wetherill Park N S W 2164 Australia P O Box 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 02 9604 3911 International 61 2 96...

Страница 2: ...ere is a Possibility of personal injury Caution alerts apply when there is a Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT BOXES located thr...

Страница 3: ...t for damage Retain the packaging materials until the equipment has been thoroughly tested Notify the detail of any damage without delay to your supplier or Thermoline Scientific LOCATION Place the eq...

Страница 4: ...controls enclosure Below is suspended the circulating pump housing heating element and sensor The desired operating temperature is set at the control A signal from the sensor is monitored at the contr...

Страница 5: ...mp soft cloth The element and pump housing refer to following note Note Stainless steel is under most conditions extremely resistant to corrosion This is in part due to the addition of chromium and ni...

Страница 6: ...f the power and consult a qualified technician The temperature range is 30 C to 100 C Ambient 5 C Model TU 1 The desired operating temperature is set at the control knob against the scale which is gra...

Страница 7: ...he electrical lead into a 10 Amp 240 Volt 50Hz GPO General Purpose Outlet 4 Select the desired operating temperature at the control knob Caution Contact with hot fluid vapours bath surfaces or product...

Страница 8: ...ng the preceding clause and to the extent permissible by law the liability of Thermoline is limited in relation to the warranted product and at the option of Thermoline to replacing the product or the...

Отзывы: