CAUTION: 

The range of equipment 

described here are not to be used in 
conjunction with large quantities of 
water. The introduction of water and the 
resultant vapour could result in 
premature component failure. 

 

 

 

 

 

WARNING: 

Do not use flammable 

solvents or combustible materials as fire 
or explosion may occur. This equipment  
uses components which may be a 
source of ignition.

 

 

 

FUNCTION 

 
To provide an accurate controlled temperature between a5°C to +60°C.

 

 

PRINCIPLE OF OPERATION 

 
The range of 

large

 above ambient incubators are ideally suited to large scale 

operations where volume throughput is desired. 
 
The models feature heavy duty insulation and multi glazed door(s), these combine to 
minimise heat loss, allow 

visual

 access to the product under evaluation whilst not 

disturbing the operating conditions. The door(s) close on a magnetic gasket and are 
further secured with positive action door latch(es).

 

 
The external is powder coated steel. 
 
The desired operating temperature is set at the temperature controller. A signal from 
the temperature sensor is monitored at the controller. The resultant calculated pulsed 
output is directed to a solid state relay which switches power to a solid state relay 
and to the ceiling mounted element(s). 
The circulating fans gently move the air across the element(s) exiting at the rear 
perforated panel, throughout the chamber before rising to the fan inlets at the ceiling. 
 
To minimise loss of conditions, the circulating fans will be isolated when the door(s) 
are open and restart upon closure.

 

 
 

PERFORMANCE 

 
The operating range for this equipment is a5°C to 60°C with a control 
accuracy of 0.2°C.

 

 
 

MAXIMUM TEMPERATURE 

 
For this product is 60

°

C.

 

 

C-TICK 

 
The equipment referred to here complies with requirement of AS/NZS 1044:1995 
electromagnetic interference (EMI). 

Содержание TI-1000F

Страница 1: ...ty Ltd T A Thermoline Scientific ABN 80 000 859 129 10 12 Ross Place Wetherill Park N S W 2164 Australia P O Box 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 02 9604 3911 Internation...

Страница 2: ...e is a Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT BOXES located throughout these operating instructions CONTENTS SAFETY INFORMATION 2 ALE...

Страница 3: ...he packaging materials until the equipment has been thoroughly tested Notify the detail of any damage without delay to your supplier or Thermoline Scientific LOCATION Place the equipment in on a firm...

Страница 4: ...ing conditions The door s close on a magnetic gasket and are further secured with positive action door latch es The external is powder coated steel The desired operating temperature is set at the temp...

Страница 5: ...eel is under most conditions extremely resistant to corrosion This is in part due to the addition of chromium and nickel to the steel and the formation of a durable chromium oxide at the surface durin...

Страница 6: ...ues to trip it may indicate an electrical fault exists Switch Off the power and consult a qualified technician TEMPERATURE CONTROLLER SENSOR This is a Eurotherm 3216 PID control with digital temperatu...

Страница 7: ...ment as fire or explosion may result This cabinet contains components that may ignite such materials WARNING 2 Do not introduce water into this equipment as the excessive vapour may result in prematur...

Страница 8: ...ntroller shifting the input value of the sensor This correction value is first entered when the equipment is under test at the factory Further small adjustments can be made by the operator to compensa...

Страница 9: ...t value with few exceptions is between 0 5 and 1 5 The need to use a value outside this range should be carefully considered If in doubt return the value to zero and recalculate the offset value The s...

Страница 10: ...able to accept the 4 20mA signals representing 0 0 to 100 0 C and 0 to 100 RH Maximum load to be applied to the analogue output is 500 ohms Pin connections for retransmission socket ALARMS No volt al...

Страница 11: ...imply contact Thermoline Scientific If warranty service is required Thermoline Scientific will endeavour to have an authorised service organisation carry out the work if there is an available agent ho...

Страница 12: ...y by Any defect in material or workmanship of or any other defect whatsoever in or unsuitability for any purpose of the goods or any part of the goods or default or negligence on the part of Thermolin...

Отзывы: