manualshive.com logo in svg
background image

 
 

Doc No.  

November 2018 

 

Thermo Scientific Dionex DRS 600, ERS 500e and ERS 500 Carbonate 

Suppressor Installation Checklist 

 

Thank you and congratulations on your recent purchase of a DIONEX suppressor product from 
Thermo  Fisher  Scientific. 

To  ensure  a  smooth  installation  experience,  please  review  and 

complete the following actions prior to installing the new suppressor: 
 
Complete  this  Installation  Checklist 

Sections  I-II

  and  check  the  box  next  to  it  to  indicate 

completion. If needed, complete 

Section III

 and check the box next to it to indicate completion. 

If  the  instruction  is  unclear  at  any  step,  refer  to  the 

Installation  Summary

  at  the  end  of  this 

document.   
 

I.

 

Backpressure Check of Detector Cell and Backpressure Coils  

Before putting the new suppressor to use, we recommend measuring the backpressure 
to  the  suppressor,  ensuring  that  the  suppressor  is  not  exposed  to  high  backpressure 
inadvertently.  Care  should  be  taken  not  to  exceed  100  psi  of  backpressure  to  the 
suppressor (150 psi for 2 mm anion versions).  The backpressure to the suppressor comes 
from any tubing, detector cell, backpressure coils, or any other components that are fluid-
connected to the eluent out port of the suppressor.   
 

Step 1:

 Disconnect the 

column out

/

eluent in

 tubing from the old suppressor and 

direct it to waste. Turn on the pump at the application flow rate for 10 minutes. 
Record the system pressure 1 (

P

1

) and turn off the pump.   

 

P

1

, ________ psi 

 

Step 2:

 Connect the 

column out/eluent in

 tubing to the detector cell 

inlet

,

 

remove 

the backpressure coils after the detector cell if installed, and direct the detector 
cell 

outlet

 tubing to waste. Turn on the pump at the application flow rate for 10 

minutes. Record the system pressure 2 (

P

2

) and turn off the pump. 

 

P

2

,

 _________psi 

 

Calculate the backpressure across the detector cell (

BP

1

):  

             

BP

1

 = 

P

2

 

 

P

1

 = ________psi                                                           

              
The  backpressure  of  the  detector  cell  is  typically 

≤ 

70  psi

  at  a  flow  rate  of  2 

mL/min.  If  the  backpressure  exceeds  this  specification,  see 

Detector  Cell 

Backflush

 Section 

 

 

                                                

 

Step  3:

  Connect  the  recommended  number  of  backpressure  coil(s)  (see  table 

below) to the detector cell 

outlet

 and direct the backpressure coil 

outlet 

to waste. 

Turn on the pump at the application flow rate for 10 minutes. Record the system 
pressure (

P

3

) and turn off the pump.  

 

 

Отзывы: