Thermo Scientific

 

     

Fresco 17 / 21 Centrifuge

Instructions for Use

50133376-

e

  •  0

6

 / 2020

Visit us online to register your warranty:

thermofisher.com/labwarranty

Содержание 75002402

Страница 1: ...Thermo Scientific Fresco 17 21 Centrifuge Instructions for Use 50133376 e 06 2020 Visit us online to register your warranty thermofisher com labwarranty...

Страница 2: ...WEEE Conformity This product is subject to the regulations of the EU Waste Electrical Electronic Equipment WEEE Directive 2012 19 EU It is marked by the following symbol...

Страница 3: ...up 2 2 Transporting the Centrifuge 2 2 Location 2 2 Aligning the Centrifuge 2 3 Mains Connection 2 3 Storage 2 4 Chapter 3 Control Panel 3 1 Control Panel 3 2 Keys 3 2 Chapter 4 Operation 4 1 Switch...

Страница 4: ...aving 5 5 Service of Thermo Fisher Scientific 5 6 Shipping and Deposing of Accessories 5 6 Chapter 6 Troubleshooting 6 1 Mechanical Emergency Door Release 6 2 Troubleshooting by User 6 3 When to conta...

Страница 5: ...ly If any parts are missing please contact your nearest Thermo Fisher Scientific representative Article number Quantity Check Centrifuge Fresco 17 21 1 Power Supply Cable 1 Rotor 1 20360104 Allen Wren...

Страница 6: ...sample density at maximum speed 1 2 Accident Prevention g cm 3 WARNING If a hazardious situation occurs turn off the power supply to the centrifuge and leave the area immediately WARNING In case of a...

Страница 7: ...r Do not use rotors which show any signs of corrosion and or cracks Do not touch the mechanical components of the rotor and do not make any changes to the mechanical components Use only with rotors wh...

Страница 8: ...ent Do not centrifuge inflammable substances Remaining risk Improper use can cause damages contamination and injuries with fatal consequences Do not centrifuge toxic or radioactive materials or any pa...

Страница 9: ...uction and Description Contents Characteristics of the Fresco 17 21 on page 1 2 Technical Data on page 1 3 Directives Standards and Guidelines on page 1 4 Functions and Features on page 1 5 Mains Supp...

Страница 10: ...d RCF or the entry mode These settings can be changed even while the centrifuge is running With the help of the key you can also centrifuge a sample for just a few seconds if called for The Fresco 17...

Страница 11: ...V 120 V 100 V 1092 BTU h 1126 BTU h 1126 BTU h IP 20 20 20 Run Time unlimited unlimited unlimited Max Speed nmax 13300 rpm 13300 rpm 13300 rpm Min Speed nmin 300 rpm 300 rpm 300 rpm Maximum RCF Value...

Страница 12: ...g 28 kg 28 kg Table 1 2 Technical Data Fresco 21 Feature Value Table 1 3 Directives Standards and Guidelines Region Marked Standard Europe 220 230 V 50 60 Hz 230 V 50 60 Hz 98 79 EC In Vitro Diagnosti...

Страница 13: ...nt data is saved Functions RCF selection temperature control pretemp Imbalance Recognition Electronic in ventilated centrifuges Door Lock Automatic door closing and locking starting from an initial ho...

Страница 14: ...x 8 Rotor 75003489 Hematocrit Rotor 75003473 Refrigerant Art No 75002420 Refrigerant R 134a Quantity 0 26 kg Pressure 21 bar GWP 1430 CO2e 0 37 t Contains fluorinated greenhouse gases in a hermetical...

Страница 15: ...c Fresco 17 21 2 1 2 Before Use Contents Before Setting up on page 2 2 Transporting the Centrifuge on page 2 2 Location on page 2 2 Aligning the Centrifuge on page 2 3 Mains Connection on page 2 3 Sto...

Страница 16: ...l people Always lift the centrifuge at both sides on the bottom plates Figure 2 1 Lifting the centrifuge at both sides The centrifuge can get damaged due to impacts Location The centrifuge should only...

Страница 17: ...ply switch on the back press 0 2 Plug the centrifuge into grounded electrical sockets only 3 Check whether the cable complies with the safety standards of your country 4 Make sure that the voltage and...

Страница 18: ...ntific Storage Before storing the centrifuge and the accessories it must be cleaned and if necessary disinfected and decontaminated Store the centrifuge in a clean dust free location Be sure to place...

Страница 19: ...Thermo Scientific Fresco 17 21 3 1 3 Control Panel Contents Control Panel on page 3 2 Keys on page 3 2...

Страница 20: ...see Mechanical Emergency Door Release on page 6 2 Pulse By pressing the key the centrifuge starts immediately and accelerates up to the end speed Releasing the key initiates a stopping process at the...

Страница 21: ...ning on page 4 2 Close Door on page 4 2 Rotor Installation on page 4 2 Entering Parameters on page 4 4 Centrifugation on page 4 8 Short term Centrifugation on page 4 8 Removing the Rotor on page 4 9 A...

Страница 22: ...n The approved rotors for the Fresco 17 21 are listed in section Rotor Selection on page 1 6 Use only the rotors and accessories from this list in the centrifuge Proceed as follows 1 Open the door of...

Страница 23: ...ockwise Hold the rotor with the other hand into position 5 Close the rotor 6 Check if the rotor is properly installed by lifting it slightly on the handle CAUTION Do not install the rotor when the tem...

Страница 24: ...s WARNING Check for any damage to the rotor Damaged rotors must not be used Keep the centrifuge spindle area of the rotor clear of objects CAUTION Check that the rotor is properly locked on the centri...

Страница 25: ...minimum of 300 rpm The maximum speed depends on the centrifuge variant Note The RCF value can be set to a minimum of 100 x g The maximum speed depends on the centrifuge variant The displayed RCF valu...

Страница 26: ...fuge will continue running until you stop it manually with the key Pre selecting the Temperature You can determine the sample temperature in C Proceed as follows 1 Enter the desired value by pressing...

Страница 27: ...the desired value shows You increase or decrease the temperature in steps of 1 C 3 Press the key The rotor will be operated at optimal speed Changing the settings during the run You can change the se...

Страница 28: ...Program With preset Run Time If the run time has been pre selected and all you have to do is to wait until the centrifuge terminates the run automatically As soon as the speed drops to zero the messag...

Страница 29: ...osol tight Rotors When using an aerosol tight lid the rotor can be removed with the lid closed This is to protect you and the samples Audible Alarm Error Accompanying all error messages a warning sign...

Страница 30: ...rned on or off 5 Press the key to confirm the pre selected value Turning off the Centrifuge 1 To turn off the centrifuge put the mains switch to 0 Note The centrifuge is equipped with a special switch...

Страница 31: ...co 17 21 5 1 5 Maintenance and Care Contents Cleaning Intervals on page 5 2 Cleaning on page 5 2 Disinfection on page 5 4 Decontamination on page 5 5 Autoclaving on page 5 5 Service of Thermo Fisher S...

Страница 32: ...since higher temperatures could damage the material and shorten the lifetime of the parts Use only disinfectants with a pH of 6 8 Dry aluminum parts off with a soft cloth After cleaning treat the enti...

Страница 33: ...After some applications there might be ice in the rotor chamber Let the ice melt and drain it off Clean the rotor chamber as described above Cleaning Filter Unit Refrigerated centrifuges have a filte...

Страница 34: ...disinfectant Adhere strictly to the given application times 7 Be sure the disinfectant can drain off the rotor 8 Rinse the rotor and accessories thoroughly with water 9 Dispose of the disinfectant acc...

Страница 35: ...f the rotors and accessories after cleaning with a cloth or in a warm air cabinet at a maximum temperature of 50 C 11 After cleaning treat the entire surface of aluminum parts with corrosion protectio...

Страница 36: ...otor and the drive shaft Thermo Fisher Scientific offers inspection and service contracts for this work Any necessary repairs are performed for free during the warranty period and afterwards for a cha...

Страница 37: ...Thermo Scientific Fresco 17 21 6 1 6 Troubleshooting Contents Mechanical Emergency Door Release on page 6 2 Troubleshooting by User on page 6 3 When to contact Service Technician on page 6 5...

Страница 38: ...braking process lasts much longer than usual Proceed as follows 1 Make sure the rotor has stopped view port in the lid 2 Pull out the power supply plug 3 Insert a 3 inch long wire e g a staple into th...

Страница 39: ...2 If this is unsuccessful you may open the centrifuge door using the mechanical emergency door release Mechanical Emergency Door Release on page 6 2 Exceptionally running noise Imbalance 1 Stop the ce...

Страница 40: ...inform Customer Service E 29 Motor does not start Motor or rotor blocked 1 Switch instrument off and on again using the mains switch 2 Open the centrifuge door 3 Check whether the rotor can turn free...

Страница 41: ...e operated Internal program error Switch the instrument off and on again If an error message appears again inform Customer Service E 60 Rotor stops with deceleration Insufficient temperature in the re...

Страница 42: ...6 6 Fresco 17 21 Thermo Scientific 6 Troubleshooting...

Страница 43: ...uminum Chloride U U S S S S U S S S S M S S S S S S S S S M U U S S Formic Acid 100 S M U U U S M U U S S U S U S U Ammonium Acetate S S U S S S S S S S S S S U S S S S S S S S S S Ammonium Carbonate...

Страница 44: ...iethylpyrocarbonate S S U S S S S S U S U S U S S S M S S S S S S Dimethylsulfoxide S S U U S S S S U S S U S U U S S U U S S S S U U Dioxane M S U U S S M M S U U S U M U U M M M U S S S S U U Ferric...

Страница 45: ...U U U S M M U U U U Potassium Permanganate S S S S S S S S S U S S S M S M S U S S M S U S Calcium Chloride M U S S S S S S S S S S S S M S S S S S S S M S S S Calcium Hypochlorite M U S M M S M S S...

Страница 46: ...S S S Oils Petroleum S S S S U S S S S U U M S M U U S S S U S S S S Oils Other S S S M S S S S U S S S S U S S S S S S M S Oleic Acid S U S S S U U S U S S M S S S S S S S S S M U S M M Oxalic Acid U...

Страница 47: ...M Trichloroacetic Acid U U U S S U M S U S U U S M M S S U U S U U U M U Trichloroethane S U M U U S U U U U U U U U U S U S S Trichloroethylene U U U U S U U U U U U U U U S U U S Trisodium Phosphate...

Страница 48: ...erials under the stress of centrifugation When in doubt we recommend pretesting sample lots Key S Satisfactory M Moderate attack may be satisfactory for use in centrifuge depending on length of exposu...

Страница 49: ...ation 5 5 Deposing 5 6 Description 1 1 Directives 1 4 Disinfection 5 4 E Emergency Door Release 6 2 F Features 1 5 Filter Unit 5 3 Functions 1 5 G Guidelines 1 4 I Intended Use iv Introduction 1 1 K K...

Страница 50: ...Index I 2 Fresco 17 21 Thermo Scientific T Technical Data 1 3 Transporting 2 2 Troubleshooting 6 1 6 3 Turning off 4 10...

Страница 51: ......

Страница 52: ...e consult your local sales represen tative for details Shown pictures within the manual are examples and may differ considering the set parameters and language Pictures of the user interface within th...