1
Introduction and Description
Directives, Standards and Guidelines
1-4
Fresco 17 / 21
Thermo Scientific
Directives, Standards and Guidelines
Overvoltage Category
II
II
II
Pollution Degree
2
2
2
Heat Dissipation
Fresco 21
Fresco 21
Fresco 21
230 V
120 V
100 V
1262
BTU/h
1297
BTU/h
1228
BTU/h
IP
20
20
20
Run Time
unlimited
unlimited
unlimited
Max Speed n
max
14800 rpm
14800 rpm
14800 rpm
Min Speed n
min
300 rpm
300 rpm
300 rpm
Maximum RCF Value at n
max
21100 x g
21100 x g
21100 x g
Maximum Kinetic Energy
<2
.
35 kNm
<2
.
35 kNm
<2
.
35 kNm
Noise Level at maximum Speed
< 50 dB (A)
< 50 dB (A)
< 50 dB (A)
Dimensions
H
e
ight
Width
Depth
3
30 mm
292
mm
440
mm
330 mm
292 mm
440 mm
330 mm
292 mm
440 mm
Weight without Rotor
28
kg
28 kg
28 kg
Table 1-2.
Technical Data Fresco 21
Feature
Value
Table 1-3.
Directives, Standards and Guidelines
Region
Marked
Standard
Europe
220-230 V, 50 / 60 Hz
230 V, 50 / 60 Hz
98/79/EC In Vitro Diagnostics
2011/65/EU RoHS
Directive on the Restriction of the use of certain Hazardous
Substances in electrical and electronic equipment
Protective goals of:
2006/42/EC Machinery
2014/35/EU Low Voltage
2014/30/EU Electromagnetic Compatibility (EMC)
EN 61010-1
EN 61010-2-020
EN 61010-2-101
EN 61326-1 Class B
EN 61326-2-6
EN ISO 14971
EN ISO 13485
USA & Canada
220-230 V, 50 / 60 Hz
120 V, 60 Hz
ANSI/UL 61010-1
UL 61010-2-020
UL 61010-2-101
EN ISO 14971
EN ISO 13485
FDA listed
Product code JQC
centrifuges for clinical use
Device class 1
Japan
100 V, 50 / 60 Hz
China
230 V, 50 / 60 Hz
208-240 V, 50 / 60 Hz
IEC 61010-1
IEC 61010-2-020
IEC 61010-2-101
IEC 61326-2-6
IEC 61326-1 Class B
EN ISO 14971
EN ISO 13485
CFDA listed