![Thermo EPD Mk2 Скачать руководство пользователя страница 85](http://html.mh-extra.com/html/thermo/epd-mk2/epd-mk2_technical-handbook_1104222085.webp)
EPD/HB/40521/000
ISSUE 5
EPD TECHNICAL HANDBOOK
Page 75
Code
Text as displayed by
EasyEPD2
EPD Action
Recommended
Customer
Action
Remarks
090
Hard Gamma detector fail*
None- continues
normal processing
Reset unit.
Return to
supplier if
recurs.
Only the first
detector to fail is
logged.
091
Soft Gamma detector fail*
None- continues
normal processing
-ditto-
-ditto-
092
Full Beta detector fail*
None- continues
normal processing
-ditto-
-ditto-
093
Beta Compensating
detector fail*
None- continues
normal processing
-ditto-
-ditto-
094
CRC error - Dose Alarm
Thres holds
Logs fault and
performs reset without
doing log first. This
means that good data
is read back from the
EEPROM. For the
ADS and Dose data,
the QF bits are also
set. This is done by
reading the data from
EEPROM, setting the
bit, calculating the
new checksum,
restoring the data to
EEPROM and then
doing a reset. The
correct values will
then be restored on
program restart.
Reset unit.
Return to
supplier if
recurs.
Indicates
corruption of data
in RAM. Up to 15
mins of dose
could be lost.
095
CRC error - Rate Alarm
Thresholds
-ditto-
-ditto-
-ditto-
096
CRC error - ADS Data
-ditto-
-ditto-
-ditto-
097
CRC error - Calibration
Factors
-ditto-
-ditto-
-ditto-
098
CRC error - Dose Data
-ditto-
-ditto-
-ditto-
* only the first detector to fail is logged
IRDA Module
Code
Text as displayed by
EasyEPD2
EPD
Action
Recommended
Customer Action
Remarks
100
IRDA Module Error
Reserved for future use
Table 5.2
EPD Faults Logged to History Event Store.
Содержание EPD Mk2
Страница 6: ......
Страница 10: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page xii ...
Страница 14: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 4 This page is intentionally blank ...
Страница 25: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 15 ...
Страница 51: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 41 This page intentionally blank ...
Страница 52: ......
Страница 57: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 47 Figure 4 1 EPD Dimensions ...
Страница 86: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 76 This page is intentionally blank ...
Страница 103: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 93 This page is intentionally blank ...
Страница 104: ......
Страница 109: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 99 APPENDIX D EPD FAULT LOG FORM SEE OVER ...
Страница 110: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 100 This page intentionally blank ...
Страница 112: ...EPD FAULT LOG FORM HB40521 ...