![Thermo EPD Mk2 Скачать руководство пользователя страница 32](http://html.mh-extra.com/html/thermo/epd-mk2/epd-mk2_technical-handbook_1104222032.webp)
EPD/HB/40521/000
ISSUE 5
EPD TECHNICAL HANDBOOK
Page 22
3.2.5
Alarm Muting
Alarm conditions cause the sounder to be activated and the alarm LED to illuminate,
subject to these facilities being enabled for the alarm. The appropriate LCD alarm
flag is always activated. The user can silence ( mute ) the alarm by pressing and
holding the button, provided mute is enabled on the alarm. This will also extinguish
the Alarm LED. However the LCD alarm flag remains set until such time as the
alarm condition is cleared.
Dose rate alarms are self muting when the dose rate falls below the alarm ‘off’
threshold level. Similarly dose alarms are self muting if the dose is cleared or the
thresholds suitably increased over the IR communications link.
Note that
if
the display backlight is enabled a short press of the button turns the back-
light on and temporarily silences any active alarm. This is because the EPD unit
battery cannot supply sufficient current for both backlight and sounder together. After
the display timeout period the backlight is automatically turned off and the sounder
re-enabled. Thus a short press may appear to mute the alarm and confuse the user
under such conditions. It is necessary therefore for the user to remember that a long
press is required to mute the alarm.
Содержание EPD Mk2
Страница 6: ......
Страница 10: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page xii ...
Страница 14: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 4 This page is intentionally blank ...
Страница 25: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 15 ...
Страница 51: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 41 This page intentionally blank ...
Страница 52: ......
Страница 57: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 47 Figure 4 1 EPD Dimensions ...
Страница 86: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 76 This page is intentionally blank ...
Страница 103: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 93 This page is intentionally blank ...
Страница 104: ......
Страница 109: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 99 APPENDIX D EPD FAULT LOG FORM SEE OVER ...
Страница 110: ...EPD HB 40521 000 ISSUE 5 EPD TECHNICAL HANDBOOK Page 100 This page intentionally blank ...
Страница 112: ...EPD FAULT LOG FORM HB40521 ...