Kelvitron® temperature controller
Start - Up ( • = Input, IP- = Display )
Turn on main switch of the unit.
Green pilot lamp lights up.
Digital displays light up for approx. 10 sec. (test phase),
After the test phase the display will indicate the actual temperature value, controller output indi-
cation shows the status of the output: active / non active (pilot lamp).
Press "C" key.
The actual set point appears an display (flashing).
Change set point with "+" or " key .
If the "+" or "-“ key will be pressed longer, the value of set point in the display runs
faster:
Set points up to 99.9 °C are in 0,1 °C steps adjustable.
Set points up to 300 °C are in 1 °C steps adjustable.
Press "
°
C" key once more.
The desired set point will be stored, the display does not flash, the display shows the actual
temperature value.
The unit heats up to the desired temperature.
If the "°C" key will not be actuated, the old set point is maintained.
Light-emitting diode "heating"
Up / down keys
Setpoint-
key"
Temperature-
display
Содержание VT 6060 M-BL
Страница 1: ...USA GB Valid 12 2015 50041981 J Vacuum Drying Oven VT 6060 M BL VT 6130 M BL Operating instructions...
Страница 4: ...4 45 50041981 J...
Страница 9: ...50041981 J 9 45...
Страница 10: ...10 45 50041981 J...
Страница 16: ...It is electrically and functionally independent of the temperature control system 16 45 50041981 J...
Страница 25: ...50041981 J 25 45...
Страница 27: ...as the overtemperature protection device 50041981 J 27 45...
Страница 30: ...30 45 50041981 J...
Страница 32: ...with small flange KF DN 25 onto the extraction 32 45 50041981 J...
Страница 33: ...50041981 J 33 45...
Страница 40: ...40 45 50041981 J...
Страница 41: ...50041981 J 41 45...