Functional Description
Mass Spectrometer
Thermo Scientific
TSQ Quantum XLS Series User Guide
31
Mass Analyzer
The mass analyzer separates ions according to their mass-to-charge ratio and then passes them
to the ion detection system. The mass analyzers on the TSQ Quantum XLS Series mass
spectrometers consist of three quadrupole rod assemblies (Q1, Q2, and Q3) and three lens
sets.
The mass analyzer is discussed in detail in the following subtopics:
•
•
Rf and dc Fields Applied to the Quadrupoles
•
•
Collision Cell and CID Efficiency
•
•
Quadrupole Rod Assemblies
The three rod assemblies used in the mass spectrometer are numbered from the ion source end
of the manifold and are designated Q1, Q2, and Q3. Q1 and Q3 are quadrupoles that enable
high-resolution scans without signal loss.
Q1 and Q3 are square arrays of precision-machined and precision-aligned, true hyperbolic
quadrupoles, called hyperquads (TSQ Quantum XLS Ultra mass spectrometer) or
hyperbolic-profile round rods (TSQ Quantum XLS mass spectrometer). See
and
. Quartz spacers act as electrical insulators between adjacent rods.
Figure 20.
Q1 or Q3 hyperquad of the TSQ Quantum XLS Ultra mass spectrometer
Figure 21.
Q1 or Q3 quadrupole of the TSQ Quantum XLS mass spectrometer
Содержание TSQ Quantum XLS
Страница 1: ...TSQ Series TSQ Quantum XLS Series User Guide 70111 97225 Revision A December 2011 ...
Страница 3: ......
Страница 4: ......
Страница 12: ......
Страница 16: ......
Страница 32: ......
Страница 64: ......
Страница 76: ......
Страница 102: ...86 TSQ Quantum XLS Series User Guide Thermo Scientific Figure 50 Ion source assembly assembled and exploded views ...
Страница 160: ......
Страница 166: ......
Страница 176: ......