6
Maintenance
Cleaning Ion Source Components
Thermo Scientific
TSQ Quantum XLS and TSQ Quantum GC User Guide
85
Cleaning Ion Source Components
An important part of maintaining your TSQ Quantum XLS or TSQ Quantum GC mass
spectrometer is making sure that the ion source components are clean. Follow the cleaning
procedures in this section to clean stainless steel and non-stainless steel parts.
When your mass spectrometer is clean and in good working order, perform benchmark tests
and record the results. When the tested performance of your system decreases significantly
from your benchmark test results, clean the ion volume. If this does not restore performance,
clean the lenses.
How often you clean the ion source depends on the types and amounts of samples and
solvents you introduce into the system. In general, the closer a component is to the sample
introduction point, the more rapidly the component becomes dirty. For example, you clean
the ion volume more often than other parts. If you just want to change or clean the ion
volumes, you do not need to shut down the system.
Most parts can be removed and disassembled by hand. Make sure you have all the tools
needed
before starting each procedure.
shows an exploded view of the ion source assembly.
Contents
•
Cleaning Ion Source Components
•
Replacing the Filament of the TSQ Quantum XLS Ion Source
•
Replacing the Filament of the TSQ Quantum GC Ion Source
•
Disassembling and Reassembling the TSQ Quantum XLS Ion Source
Completely
•
Disassembling and Reassembling the TSQ Quantum GC Ion Source
Completely
•
•
•
•