6
Maintenance
Removing and Installing a GC Capillary Column
122
TSQ Quantum XLS and TSQ Quantum GC User Guide
Thermo Scientific
Removing and Installing a GC Capillary Column
You replace the GC capillary column if the chromatographic separation is bad or if it does not
have the appropriate stationary phase for your application.
Removing a GC Capillary Column
Y
To remove a GC capillary column
1. Shut down the TSQ Quantum XLS or TSQ Quantum GC and the gas chromatograph.
a. Shut down and vent the mass spectrometer. See
b. Lower the oven, injector, and transfer line temperatures to 30 °C and allow them to
cool before continuing.
c.
Once the oven, injector, and transfer line are cool, turn off the gas chromatograph.
2. Remove the column from the transfer line. See
.
a. Unscrew the transfer line nut.
b. Remove the column from the transfer line.
Tools Needed
• Gloves, cleanroom grade (P/N 23827-0008 and 23827-0009)
• Wrench, open-ended, 5/16 in.
• Wrench, open-ended, 7/16 in.
• Wrench, open-ended, 6 mm
Frequency
As needed for maintenance or column replacement
CAUTION
The oven, injector, and transfer line are hot. Allow them to cool to
room temperature before touching them.