Getting Familiar with Your TriPlus RSH
ITEX-2 Tool
Thermo Scientific
TriPlus RSH User Guide
31
Besides the coupling pivots for the needle guide and the tool, the purge gas connector is
directly installed on the cover plate of the tool. The syringe flange is supported by the polymer
counterpart to the retaining nut. In the heating cylinder there are two O-rings which seal the
gas flow to enable a syringe purge. It is recommended to replace these O-rings during
preventative maintenance.
shows the cut-out of the ITEX tool visualizing the Purge Gas Flow and the Sealing
O-ring.
Figure 19.
ITEX Tool Cut-out (1)
The heating and cooling functionality for the Trap is integrated in the lower part of the tool.
See
Figure 20.
ITEX Tool Cut-out (2)
del gas di lavaggio
WARNING - BURN HAZARD
The ITEX Tool can operate at a maximum temperature of
200 °C. Operating at this temperature, its outer surface will reach about 70 °C. Do not
touch the outside of the ITEX Tool. Always let the ITEX Tool cool down before you
touch it, or before you remove the Syringe.
Trap Cooler
Trap Heater