5
Service Menu Item
Installation Item
Thermo Scientific
TriPlus RSH User Guide
213
4. When finished, the final wizard step moves the Head back to the Home position.
#
Setup Module
This allows you to predefine the parameters from a module. Typically, this process is done
once at first installation of the TriPlus RSH. See
.
Figure 156.
Wizard Item Setup Module
To setup a module
The example below shows the steps for defining the parameters for the
Agitator
:
CAUTION
TR Station or ATC Station should not change in the position. If the Station
has moved its position, it is necessary to redefine the Reference Point, and recalibrate all
the other tools.
CAUTION
If for any reason, the referencing (Set Reference Point) has been repeated, it is
mandatory to recalibrate all the tools. Using the step
Set Reference Point
clears the
calibration data for the individual tools.