TriPlus RSH Site Preparation
Power Requirements
Thermo Scientific
TriPlus RSH Preinstallation Requirements Guide
9
Electrical Specifications
The electrical specifications and the various protection classes are provided belo
Use the following guidelines to ensure your site is equipped with enough power to support the
system.
CAUTION
The Power Module should be placed on a sturdy, level bench with adequate
access to the main power switch.
Table 6.
Electrical Specifications (Sheet 1 of 2)
Parameter
Requirements
Protection Class
Class I
Over voltage category
Category II
Pollution degree
2
Moisture protection
Normal (IPX0)
Voltage 36
VDC
Current
3.2 A
Fuse
T6.3 A/250 V
Power Supply; Handheld Controller
Requirements
Input line voltage
Grounded AC, 100 to 240 V
Input line frequency
50/60 Hz
Input power
5 A
Output Voltage
36 VDC
Output current
5.55 A (total for two Outlets)
Table 7.
System Power Requirements (Sheet 1 of 2)
Equipment
Maximum Power (W)
TriPlus RSH
200 W with a single power module
400 W with two power modules
TRACE GC Ultra
1920
FOCUS GC
1600
TSQ 8000 Series MS, including foreline
pump
1080
ISQ Series MS, including foreline pump
1200