11
Upgrade Equipment
Upgrading a Stand Alone TRACE 1300/TRACE 1310 to MS Version
574
TRACE 1300 and TRACE 1310 Hardware Manual
Thermo Scientific
f.
On the right side wall in the interior of the oven, remove the partial cut plate from
the corresponding duct. See
.
Figure 636.
Perform the Duct for TSQ Quantum (4)
4. Replace the right side panel with the right panel provided for MS.
a.
Using a T20 Torxhead screwdriver, unscrew the duct plate screws from the right
panel for MS. See
.
Figure 637.
Right Panel for MS
b. Plug the ground wire previously removed to the left panel of MS provided.
c.
Place the right panel for MS and attach the screw holding it in place.
5. Introduce the TSQ Quantum transfer line inner tube into the oven through the duct
provided.
6. Attach the transfer line to the GC column using the proper nut and ferrule.
7. Place the Aux Temperature/Cryo module into the housing on the back of the GC.
See the section
“Adding an Aux Temperature/Cryo Module”
for details.
Note
T
he column must be conditioned before installing into the transfer line.
“Installing the Column the First Time”
.
Duct for TSQ Quantum Partial Cut Plate
Duct Plate