3
Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
109
Figure 97.
HeS-S/SL Module Components
Maintaining the Instant Connect Helium Saver Injector Module is largely the same as a
conventional SSL inlet.
One advantage of the Instant Connect Helium Saver Injector Module
is that routine septum
and liner changes can be accomplished without cooling the MS transfer line or ion source.
This is particularly desirable since cooling down and re-establishing stable MS temperatures
takes much longer than cooling down and re-establishing the inlet temperature.
The Instant Connect Helium Saver Injector Module periodic maintenance includes:
• Replacing the septum
The septum needs to be changed intermittently to prevent leakage. Replace the septum at
least after every 200 injections, or every time a problem related to septum damage, or
wear occurs.
See
.
• Cleaning or replacing the liner
Injection port liner needs to be replaced or cleaned as it becomes dirty.
See
“Cleaning or Replacing the Glass Liner”
Septum Cap
Septum
Ring Nut
Septum Holder/Liner Cap
Liner Seal
Liner
Body Head External O-Ring
Body Head Internal O-Ring
Bottom Fittings
Tip
It is good practice to replace the septum every time you replace the glass liner.