3
Performing Injectors Routine Maintenance
Maintaining an Instant Connect Helium Saver Injector (HeS-S/SL)
Thermo Scientific
TRACE 1300 and TRACE 1310 Hardware Manual
111
Replacing the Septum
To replace the septum
Figure 98.
HeS-S/SL Injector: Septum Replacement
1. Put the GC in standby condition.
2. Cool the oven and injector to room temperature.
3. Ensure the
E
nable conservation
field in the configuration page of the inlet reads
Yes
.
4. Check that there is at least 20 mL/min split flow exiting the split line, then turn the
column flow to
Off
in the GC user interface and allow the inlet to depressurize.
5. Put the autosampler away if present.
6. Open the module flap cover.
7. Replace the septum.
a.
Unscrew and remove the septum cap.
b. Using tweezers, remove the septum from the septum holder.
c.
Avoid touching the septum with your fingers. Insert a new septum into the septum
holder using tweezers.
d. Screw and tighten the septum cap to finger-tight.
Note
Care should be taken when performing this procedure in order to keep from
damaging analytical columns.
Note
By pressing the
Maintenance
button, the GC cool down is automatically
carried out.
Septum Holder
Septum
Septum Cap