Dionex SRD-10 Installation Instructions
Doc. 065355-04 05/18
5
c. Connect the wires from the Dionex SRD-10 connector plug to the TTL
connector pins on the other device.
NOTE Check the polarity of the connection. Connect the
signal wire to the signal (+) pin and the ground wire
to the ground (-) pin.
8. Plug the USB cable from the Dionex SRD-10 into any available USB port on
the IC system (or the optional external USB power supply).
9. Turn on the power to the IC system.
3.
Chromeleon Setup
Before beginning operation of the Dionex SRD-10, you must verify that a pump
TTL input is configured in Chromeleon to control the eluent pump. In addition,
you may need to update the pump firmware. The following table indicates the
setup requirements for each pump.
The setup instructions vary slightly, depending on the chromatography software
used to control your IC system. If you are running Chromeleon 7, refer to
. If you are running Chromeleon 6.8, refer to
Product
Update pump firmware
to version 4.1.0 (or
later)?
Configure pump TTL
input to control eluent
pump?
Dionex ICS-1000, ICS-1100,
ICS-1500, ICS-1600, ICS-2000,
or ICS-2100
No
Yes
Dionex ICS-3000 Dual Pump
(DP) or ICS-3000 Single Pump
(SP)
Yes
Yes
Dionex ICS-5000 Dual Pump
(DP) or ICS-5000 Single Pump
(SP)
No
Yes