![Thermo Scientific Thermo Scientific Dionex SRD-10 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/thermo-scientific/thermo-scientific-dionex-srd-10/thermo-scientific-dionex-srd-10_installation-instructions-manual_1105376010.webp)
Dionex SRD-10 Installation Instructions
8
Doc. 065355-04 05/18
b. Insert the Dionex ICS-3000 Dual Pump/ICS-3000 Single Pump
Firmware CD into the appropriate drive.
c. Copy the pump firmware file from the CD to the
\Bin
directory of
your Chromeleon installation.
3. Click
Install Firmware
again. When the notification of the installed
firmware version appears, click
Yes
to update the firmware.
4. The download can take several seconds. When the download is
complete, remove the firmware CD from the drive. Store the CD in a
safe place, away from heat and moisture.
Configure the Pump TTLs
1. Click the
Inputs
tab in the pump Properties dialog box. Verify that
the correct TTL is selected.
For a Dionex ICS-3000 DP or ICS-5000 DP:
If the Dionex SRD-10
is connected to Pump 1 and TTL 2, do
not
select the
Pump_TTL_2
check box. If the Dionex SRD-10 is connected to Pump 2 and TTL 3,
do
not
select the
Pump_TTL_3
check box.