Thermo Scientific
Model 3900 Series |
4-1
Installation |
Chapter 4
Installation
Ambient Conditions
Requirements
The environmental chamber must only be operated in a location that meets all of the
ambient condition requirements listed below:
Installation location indoors in dry areas free from drafts.
The following minimum distances must be observed: lateral 150 mm, front 1100
mm, back 250 mm and top 250 mm.
The dust pollution may not exceed the contamination category 2 based on EN
61010-1. Using the environmental chamber in an atmosphere with electrically
conductive dust is prohibited.
The room must be equipped with appropriate ventilation. Solid, level, fire-proof
surface; no flammable materials opposite to the rear panel of the environmental
chamber.
The electrical circuitry of the environmental chamber has been designed for an
operating height of up to 2000 m above sea level.
If a high-voltage test is to be performed on the unit, it must first be heated for around
30 minutes at 50°C.
The ambient temperature must be within 18°C - 26°C (64°F - 79°F) with light
module and 16°C - 32°C (61°F - 90°F) without light module, see “Specifications”.
Devices that produce excessive amounts of heat must not be placed near the
environmental chamber.
Power line voltage variations must not exceed ±10 % of the nominal voltage.
Transient surges must lie within the range of levels that normally occur in the power
supply system. The impulse withstand voltage based on surge category II of IEC
60364-4-443 shall be applied at the nominal voltage level.
Relative humidity up to 70%, non condensing, at an ambient temperature of 28°C,
decreasing linearly up to 55% at 32°C without light module, up to 70%, at 28°C with
light module.
Please note: Should condensation exist, wait until the moisture has evaporated
completely before connecting the environmental chamber to a power source and
powering up.
Содержание Thermo Scientific 3906
Страница 6: ...Table of Contents A 4 Model 3900 Series Thermo Scientific ...
Страница 12: ... Safety Notes 1 6 Model 3900 Series Thermo Scientific Chapter 1 ...
Страница 22: ... Installation 4 6 Model 3900 Series Thermo Scientific Chapter 4 ...
Страница 42: ... Product Description 5 20 Model 3900 Series Thermo Scientific Chapter 5 ...
Страница 46: ... Operation 7 2 Model 3900 Series Thermo Scientific Chapter 7 ...
Страница 76: ... Disposal 11 2 Model 3900 Series Thermo Scientific Chapter 11 ...
Страница 84: ... Specifications 12 8 Model 3900 Series Thermo Scientific Chapter 12 ...
Страница 94: ... Service 13 10 Model 3900 Series Thermo Scientific Chapter 13 ...
Страница 95: ...Thermo Scientific Model 3900 Series 13 11 Service Chapter 13 ...
Страница 98: ... Service 13 14 Model 3900 Series Thermo Scientific Chapter 13 Honeywell Truline Configuration Record ...
Страница 99: ...Thermo Scientific Model 3900 Series 13 15 Service Chapter 13 ...
Страница 100: ... Service 13 16 Model 3900 Series Thermo Scientific Chapter 13 ...
Страница 101: ...Thermo Scientific Model 3900 Series 13 17 Service Chapter 13 ...
Страница 102: ... Service 13 18 Model 3900 Series Thermo Scientific Chapter 13 ...