Thermo Scientific
Model 3900 Series |
1-3
Safety Notes |
Chapter 1
Disconnect the unit from all power sources before cleaning, troubleshooting, or
performing other maintenance on the product or its controls. To disconnect power
supply to the environmental chamber, unplug the supply cord at the back of the
chamber. Note that deactivating the main switch on the front control panel to the Off
position is not sufficient to disconnect power.
The contents of these operating instructions are subject to change at any time
without further notice.
Keep these operating instructions close to the environmental chamber so that safety
instructions and important information are always accessible.
Should you encounter problems that are not detailed adequately in these operating
instructions, please contact Thermo Fisher Scientific immediately for your own.
Operational Safety Rules
The following rules must be heeded when working with environmental chambers:
Observe the sample weight limits specified for your environmental chambers, whole
and its shelving in particular; see “Specifications” on page 12-1.
Arrange the samples evenly throughout the work space, making sure not to place
them too closely to the interior walls to ensure a uniform temperature distribution.
Do not load your environmental chambers with substances that exceed the
capabilities of the available lab apparatus and personal protection equipment to
provide sufficient degrees of protection to users and third parties.
Check the door seal every 12 months for proper sealing performance and possible
damage.
Do not process any samples containing hazardous chemical substances that may be
released into the ambient air through defective seals or may cause corrosion or other
defects on parts of the environmental chamber.
The tempering of defined substances or materials with higher moisture content can
result in increased condensation in the chamber. Measures must be observed.
The user is responsible for carrying out appropriate decontamination procedures
when hazardous materials are spilled on or inside the environmental chamber.
Humidity
After transport and decommissioning, or storage under humid conditions a
drying-out process must be performed. During the drying-out process the
equipment cannot be assumed to meet all the safety requirements of the IEC
61010-2-010 standard. The drying-out period is 2 hours.
If the environmental chamber is not used in the manner specified in this
operating manual, the protection provided by the equipment design maybe
impaired.
Содержание Thermo Scientific 3906
Страница 6: ...Table of Contents A 4 Model 3900 Series Thermo Scientific ...
Страница 12: ... Safety Notes 1 6 Model 3900 Series Thermo Scientific Chapter 1 ...
Страница 22: ... Installation 4 6 Model 3900 Series Thermo Scientific Chapter 4 ...
Страница 42: ... Product Description 5 20 Model 3900 Series Thermo Scientific Chapter 5 ...
Страница 46: ... Operation 7 2 Model 3900 Series Thermo Scientific Chapter 7 ...
Страница 76: ... Disposal 11 2 Model 3900 Series Thermo Scientific Chapter 11 ...
Страница 84: ... Specifications 12 8 Model 3900 Series Thermo Scientific Chapter 12 ...
Страница 94: ... Service 13 10 Model 3900 Series Thermo Scientific Chapter 13 ...
Страница 95: ...Thermo Scientific Model 3900 Series 13 11 Service Chapter 13 ...
Страница 98: ... Service 13 14 Model 3900 Series Thermo Scientific Chapter 13 Honeywell Truline Configuration Record ...
Страница 99: ...Thermo Scientific Model 3900 Series 13 15 Service Chapter 13 ...
Страница 100: ... Service 13 16 Model 3900 Series Thermo Scientific Chapter 13 ...
Страница 101: ...Thermo Scientific Model 3900 Series 13 17 Service Chapter 13 ...
Страница 102: ... Service 13 18 Model 3900 Series Thermo Scientific Chapter 13 ...