Daily Operation
To optimize the performance of your Surveyor MSQ Plus MS detector, you need to perform
various routine operations both before and after you operate the system.
Things to Do Before Operating the MS Detector
This section contains the following procedures that you might want to perform on a daily
basis before you begin your analyses:
•
•
•
Checking the Oil Level in the Oil Mist Filter
Checking the Nitrogen Supply
Check the nitrogen supply on the regulator of the nitrogen gas tank or liquid nitrogen boil-off
tank. Make sure that you have sufficient gas for your analysis. Typical nitrogen consumption
is 15800 liters per day (based on a 24 hr day). If necessary, replace the tank. Verify that the
pressure of nitrogen reaching the mass spectrometer is between 520 ± 35 kPa (75 ± 5 psi).
If necessary, adjust the pressure with the tank pressure regulator.