System Shutdown, Startup, and Reset
Placing the System in the Off Mode
98
Surveyor MSQ Plus Hardware Manual
Thermo Scientific
Turning the MS Detector Off from Xcalibur
Y
To turn off the MS detector from the Info View - Status page of Xcalibur
1. If you are using the optional cone wash pump, turn it off by turning its power switch to
the Off position.
2. If Xcalibur is not open, open it by double-clicking the
Xcalibur
icon on the Windows
desktop.
3. If the Info view is not displayed, choose
View > Info View
to display it. Then, click the
Status tab to open the Status page.
4. To turn off the flow from the LC pump:
• If your LC pump is controlled from the Xcalibur data system, right-click the pump
listing on the Status page. Then, choose
Turn Device Into Standby
from the
shortcut menu. See
.
• If your LC pump is not controlled from the Xcalibur data system, turn it off from its
control keypad.
Figure 76.
Info View- Status page, showing the status of the Surveyor MS Pump and its
shortcut menu
5. Right-click the Surveyor MSQ listing on the Status page. Then, choose
Turn Device Off
from the shortcut menu to place the MS detector in the Off mode. See
.
The status LED on the front of the Surveyor MSQ Plus MS detector turns yellow.
Turning off the
pump flow