3
Description of the device
Workspace atmosphere
Thermo Scientific
S
TERI-CYCLE
i160
/
i250
3-9
Figure 3-5.
Air duct
The air duct consists of two shaped pieces of sheet metal:
1. Ceiling duct (1/
)
2. Rear wall duct (2/
3. Back wall of the workspace (3/
The air duct and HEPA-filter can be installed and removed without any tooling.
CO
2
supply:
To ensure the growth conditions for the cell and tissue cultures, the workspace is supplied with CO
2
.
The pH of the bicarbonate-buffered culture media largely depends on the CO
2
content of the
workspace atmosphere.
The CO
2
content of the workspace atmosphere can be controlled within a range of 0-20 %.
The supplied CO
2
must have one of the following quality characteristics:
• Purity 99.5 % min
• medical gas quality.
O
2
supply:
If the CO
2
incubator is to be operated with more than 21 % oxygen, the workspace is supplied with
oxygen (optional).
The O
2
content of the workspace atmosphere can be controlled within a range of 21 % to 90 %.
Operation with higher oxygen concentrations requires that the fire protection caution notes in
in the section
3
1
2
Содержание STERI-CYCLE i160
Страница 8: ...Table of Contents 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 12: ...Figures 4 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 76: ...5 Operation Starting operation 5 6 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 128: ...6 Handling and control Measures after power outage 6 52 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 154: ...9 Maintenance Replacing the Door Seal 9 12 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 166: ...11 Specifications STERI CYCLE i250 11 10 STERI CYCLE i160 i250 Thermo Scientific ...
Страница 196: ...13 Device log 13 2 STERI CYCLE i160 i250 Thermo Scientific ...