Orion Star A221 and Star A321 Portable pH Meters

7

pH Buffer Group Selection

The selected buffer group allows for the automatic recognition of certain pH buffers during a pH calibration. 
The USA buffer group includes pH 1.68, 4.01, 7.00, 10.01 and 12.46 buffers and the DIN buffer group includes 
pH 1.68, 4.01, 6.86, and 9.18 buffers.

1.  In the measurement mode, press 

 

2. Press 

 or 

 to highlight 

pH Channel

 and press 

f3 (select)

.

3. Press 

 

or

 

 to highlight 

Mode and Settings 

and press 

f3 (select)

.

4. Press 

 

or

 

 to highlight 

Buffer Group 

and press 

f3 (select)

.

5. Press 

 

or

 

 keys to highlight 

USA

 or

 DIN

 and press 

f3 (select)

6. Press 

 to return to the measurement mode.

Read Type Selection

1.  In the measurement mode, press 

2. Press 

  or 

 to highlight 

pH Channel

 and press 

f3 (select)

.

3. Press 

 

or

 

 to highlight 

Mode and Settings 

and press 

f3 (select)

.

4. Press 

 

or

 

 to highlight 

Read Type

 and press 

f3 (select)

.

5. Press 

 

or

 

 to highlight 

Auto

Continuous

 or 

Timed

 

(Star A321 meter only)

 and press 

f3 (select)

.

 a. 

Star A321 meter only

 – If 

Timed

 is selected and the time interval needs to be changed – highlight 

 

Timed

; press 

 to highlight hours (HH), minutes (MM) or seconds (SS); press 

f3 (edit)

 to access the 

  numeric entry screen; use the numeric entry screen to change the values and press 

f1 (back)

 when the 

  time interval is correct. 

6. Press 

 to return to the measurement mode.

Содержание Star A221

Страница 1: ...portion of the battery cover from the meter use a coin or your nger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and inser...

Страница 2: ...ad slash through it based on the last saved calibration and measurement stability M 100 Indicates a method is in use and the number of the method being used pH Indicates the type of measurement and de...

Страница 3: ...h key on the display Press to turn the meter on When the meter is on press and quickly release to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press...

Страница 4: ...it to access the numeric entry screen and edit the pH buffer value i Press or to highlight a number decimal point or negative sign press f3 enter to select the highlighted item and repeat until the bu...

Страница 5: ...ng record the pH and temperature of the sample Press again to start a new measurement b If the meter is in continuous mode the meter will immediately start taking a measurement and update the display...

Страница 6: ...numeric entry screen will popup on the display ii Press or to highlight a number decimal place or negative sign press f3 enter to select the highlighted item and repeat until the desired value is sho...

Страница 7: ...to highlight USA or DIN and press f3 select 6 Press to return to the measurement mode Read Type Selection 1 In the measurement mode press 2 Press or to highlight pH Channel and press f3 select 3 Pres...

Страница 8: ...or to highlight pH RmV ORP or Temperature as the calibration type and press f2 select 6 The meter will display a list of calibrations for the selected channel and calibration type The list shows the s...

Отзывы: