Orion Star A221 and Star A321 Portable pH Meters

5

Measurement

Press 

 while taking a measurement in the continuous measurement mode to freeze the display and 

press 

  a second time to unfreeze the display and continue the measurement. Press 

 while taking 

a measurement to manually export the measurement  to the data log, if the data log is enabled in the 
setup menu.

1.  Rinse the pH electrode and any other electrodes in use with distilled water, blot dry with a 

lint-free tissue and place into the sample.

2.  Start the measurement and wait for it to stabilize. 

a.  If the meter is in 

AUTO-READ

 mode (default setting), press 

 to start the measurement. When 

the 

 icon stops flashing, record the pH and temperature of the sample. Press 

 again to start 

a new measurement.

b.  If the meter is in continuous mode, the meter will immediately start taking a measurement and 

update the display whenever the measurement changes. Wait for the display to show 

ready

 and 

record the pH and temperature of the sample.

c. 

Star A321

 meter only -

 If the meter is in timed mode, the meter will log measurements at the 

preselected time interval, regardless of the measurement stability. The meter will update the display 
whenever the measurement changes, so the pH and temperature of the sample can be recorded 
when the display shows 

ready

.

3.  Remove the electrode from the sample, rinse with distilled water, blot dry and place into the next sample.

4.  Repeat steps 2 and 3 for all samples.

5.  When all samples have been measured, store the electrode according to its user guide.

Содержание Star A221

Страница 1: ...portion of the battery cover from the meter use a coin or your nger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and inser...

Страница 2: ...ad slash through it based on the last saved calibration and measurement stability M 100 Indicates a method is in use and the number of the method being used pH Indicates the type of measurement and de...

Страница 3: ...h key on the display Press to turn the meter on When the meter is on press and quickly release to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press...

Страница 4: ...it to access the numeric entry screen and edit the pH buffer value i Press or to highlight a number decimal point or negative sign press f3 enter to select the highlighted item and repeat until the bu...

Страница 5: ...ng record the pH and temperature of the sample Press again to start a new measurement b If the meter is in continuous mode the meter will immediately start taking a measurement and update the display...

Страница 6: ...numeric entry screen will popup on the display ii Press or to highlight a number decimal place or negative sign press f3 enter to select the highlighted item and repeat until the desired value is sho...

Страница 7: ...to highlight USA or DIN and press f3 select 6 Press to return to the measurement mode Read Type Selection 1 In the measurement mode press 2 Press or to highlight pH Channel and press f3 select 3 Pres...

Страница 8: ...or to highlight pH RmV ORP or Temperature as the calibration type and press f2 select 6 The meter will display a list of calibrations for the selected channel and calibration type The list shows the s...

Отзывы: