Orion Star A221 and Star A321 Portable pH Meters

4

pH Calibration

One to five pH buffers can be used for calibration. Always use fresh pH buffers and select buffers that bracket 
the sample pH and are one to four pH units apart. Prepare the pH electrode according to the instructions in 
the electrode use guide. Connect the pH electrode and any other electrodes to be used (ATC probe, reference 
electrode) to the meter. Power on the meter and set the measurement mode to pH.

1.  In the measurement mode, press 

f1 (cal)

.

2.  Rinse the pH electrode and any other electrodes in use with distilled water, blot dry with a lint-free tissue 

and place into the pH buffer.

3.  When the electrode and buffer are ready, press 

f3 (start)

.

4.  Wait for the pH value on the meter to stabilize and stop flashing and perform one of the following actions:

a. Press 

f2 (accept)

 to accept the displayed pH value.

b. Press 

f3 (edit)

 to access the numeric entry screen and edit the pH buffer value.

i. Press 

 or 

 to highlight a number, decimal point or negative sign; press 

f3 (enter)

 to select the highlighted item and repeat until the buffer value at the measured 

temperature is shown above the numeric entry screen. 

ii. Press 

f2 (done)

 to exit the numeric entry screen. 

iii. Press 

f2 (accept)

 to accept the entered pH value.

5. Press 

f2 (next)

 

to proceed to the next buffer and repeat steps 2 through 4 or press 

f3 (cal done)

 to save 

and end the calibration. If five buffers are used, the calibration will save and end once the fifth buffer value 
is accepted.

a.  If a one point calibration is performed, press 

f2 (accept)

 to accept the displayed slope value or press 

f3 (edit)

 to access the numeric entry screen, enter the slope value and press 

f2 (accept)

6.  The meter will display the calibration summary including the average slope. Press 

f1 (meas)

 to export 

the data to the calibration log or press 

f2 (print)

 

to export the data to the calibration log and a printer or 

computer. The meter will automatically proceed to the measurement mode.

Содержание Star A221

Страница 1: ...portion of the battery cover from the meter use a coin or your nger and release the bottom portion of the battery cover d Orientate the batteries as shown in the battery compartment housing and inser...

Страница 2: ...ad slash through it based on the last saved calibration and measurement stability M 100 Indicates a method is in use and the number of the method being used pH Indicates the type of measurement and de...

Страница 3: ...h key on the display Press to turn the meter on When the meter is on press and quickly release to turn the display backlight on or off or hold down to turn the meter off In the measurement mode press...

Страница 4: ...it to access the numeric entry screen and edit the pH buffer value i Press or to highlight a number decimal point or negative sign press f3 enter to select the highlighted item and repeat until the bu...

Страница 5: ...ng record the pH and temperature of the sample Press again to start a new measurement b If the meter is in continuous mode the meter will immediately start taking a measurement and update the display...

Страница 6: ...numeric entry screen will popup on the display ii Press or to highlight a number decimal place or negative sign press f3 enter to select the highlighted item and repeat until the desired value is sho...

Страница 7: ...to highlight USA or DIN and press f3 select 6 Press to return to the measurement mode Read Type Selection 1 In the measurement mode press 2 Press or to highlight pH Channel and press f3 select 3 Pres...

Страница 8: ...or to highlight pH RmV ORP or Temperature as the calibration type and press f2 select 6 The meter will display a list of calibrations for the selected channel and calibration type The list shows the s...

Отзывы: