Error number
Description
Troubleshooting
E-019
Rotor Unknown
Restart the centrifuge�
Check, if the rotor is qualified for the
centrifuge (check
)�
If the error message shows again, contact
the customer service�
E-025
E-027
Read Manual
Check, if the centrifuge door is blocked�
Restart the centrifuge�
If the error message shows again, contact
the customer service�
E-029
E-045
Read Manual
Check, if a rotor is installed�
Check, if the rotor is qualified for the
centrifuge (check
)�
Restart the centrifuge�
If the error message shows again, contact
the customer service�
E-030
Power Failure
Check the power supply of the centrifuge�
Make sure not to operate too many devices
at one power source�
Let the centrifuge cool down for 15 min�
If the error message shows again, contact
the customer service�
E-098
Imbalance Load
Check the load placed in the rotor�
Check that the rotor cross bolts are
greased well�
Restart the centrifuge�
If the error message shows again, contact
the customer service�
E-060
Temp Low!
CAUTION
Icy metal parts!
Restart the centrifuge� Open the centrifuge
door� Wait 120 min� Remove ice and water
out of the centrifugation chamber�
If the error message shows again, contact
the customer service�