6. Autoclaving
1�
Before autoclaving clean rotor and accessories as described above�
2�
Place the rotor on a flat surface�
Rotors and adapter can be autoclaved at 121 °C�
The maximum permissible autoclave cycle is 20 minutes at 121 °C�
NOTICE
Never exceed the permitted temperature and duration when autoclaving� No chemical
additives are permitted in the steam�
Clean the rotor before autoclaving and rinse it with distilled water� Remove all accessories (tubes,
adapters) from the rotor�
Make sure that the necessary sterility is achieved according to your requirements�
7. Maintenance
Preventive Maintenance
CAUTION
Usage beyond the rotor and lid latch lifetime limits might lead to sample
loss and damage to the centrifuge.
In order to keep this product able to perform the intended application in a reliable and safe state,
ongoing preventive maintenance is necessary in accordance with the following recommended
schedule:
For rotor lifetime refer to the specific section within the rotor user manual�
The lid latch is designed for 75 000 cycles of service�