Technical Specifications
1. 6. 5. PCR 8 x 8 Rotor
Items Supplied
Item
Article No�
Quantity
PCR 8 x 8 Rotor
75003489
1
Table 26:
Items Supplied PCR 8 x8 Rotor
General Technical Data
Maximum Permissible Load
8 x 4 g (64 x 0.5 g)
Maximum Number of Cycles
50 000
Radius (max. / min.)
7.0 cm / 4.4 cm
Angle
60°
Aerosol-tight
No
Max. Autoclaving Temperature
121 °C
Table 27:
General Technical Data PCR 8 x 8 Rotor
Rotor Performance Data
17 Series Centrifuges – PCR 8 x 8 Rotor
Voltage
230 V
120 V
Maximum Speed
13 300
13 300
Maximum RCF-Value
13 800
13 800
Acceleration / Braking Time
7 s / 8 s
7 s / 8 s
Sample Heating at max speed, 23 °C
ambient temperature, 60 min running time
31 °C
31 °C
21 Series Centrifuges – PCR 8 x 8 Rotor
Voltage
230 V
120 V
Maximum Speed
14 800
14 800
Maximum RCF-Value
17 100
17 100
Acceleration / Braking Time
8 s / 9 s
8 s / 9 s
Sample Heating at max speed, 23 °C
ambient temperature, 60 min running time
32 °C
32 °C