MaxQ 6000

Stackable Incubated and Refrigerated Shakers

Operating Manual and Parts List LT2120X1 (7004352) Rev. 11

Содержание SHKE6000 (4359)

Страница 1: ...isit us online to register your warranty www thermoscientific com labwarranty User Manual MaxQ 6000 Stackable Incubated and Refrigerated Shakers Operating Manual and Parts List LT2120X1 7004352 Rev 11...

Страница 2: ...13 Removed warranty card from page 4 1 and expanded clearance on page 4 2 ccs 7 29101 OS 567 11 30 12 New fuses connector power entry and fuse drawer for UL pgs 8 2 9 1 9 2 list ccs 6 29034 9 10 12 A...

Страница 3: ...ents and maintenance must be performed by qualified service personnel s Material in this manual is for information purposes only The contents and the product it describes are subject to change without...

Страница 4: ...that has been put on the market after 13 August 2005 This product is required to comply with the European Union s Waste Electrical Electronic Equipment WEEE Directive 2002 96 EC It is marked with the...

Страница 5: ...re or replacement parts or provide you with on site service We can also provide you with a quotation on our Extended Warranty for your Thermo Scientific products Whatever Thermo Scientific products yo...

Страница 6: ...1 Turning Shaker On 5 1 Leveling the Unit 5 1 Over Temperature Protection OTP 5 2 Setting Operating Temperature 5 2 Temperature Calibration 5 3 AC Power Loss 5 3 Setting Timer for Timed Shaking 5 4 S...

Страница 7: ...nting Plate 6 1 Cleaning Condenser Filter Coil 6 2 Platform Maintenance 6 2 Troubleshooting 7 1 Heating Error Codes 7 2 General Error Codes 7 3 Shaking Error Codes 7 3 Notes 7 4 Wiring Diagram 8 1 Rep...

Страница 8: ...lectrical outlet of correct voltage and current handling capacity 2 Disconnect from the power supply prior to maintenance and servicing To avoid personal injury 1 Do not use in the presence of flammab...

Страница 9: ...pped with the equipment or are specified for it Substituting other attachments or accessories can produce hazards or make the unit inoperative 9 Perform regular maintenance service as specified in thi...

Страница 10: ...ambient to 80 C 0 1 C at 37 C in flask SHKE6000 7 15 C below ambient to 80 C 0 1 C at 37 C in flask Minimum setpoint is 4 C on refrigerated model SHKE6000 8CE 15 C below ambient to 80 C 0 1 C at 37 C...

Страница 11: ...10 of normal rated line Temperature 15 C to 32 C Product Usage Indoor use only Refer to IEC 664 1 Caution Operating this shaker in a CO2 enriched atmosphere is not recommended The formation of carbon...

Страница 12: ...ned to provide increased aeration in a stable temperature environment Unintended Use 1 Not intended for use in Class I or II applications as defined in 21 CFR 2 Not intended for mixtures of flammable...

Страница 13: ...oad 20 lbs Maximum platform load 35 lbs All models control temperature time and speed by a Proportional Integral Derivative PID microprocessor based controller that is adjustable on a keypad in 1 rpm...

Страница 14: ...speed panel LED When chamber temperature deviates more than 1 C from setpoint an alarm will sound and the display will indicate an error 1 Speed Display 3 digit LED indicates actual or set point spee...

Страница 15: ...rs or minutes 18 Lightbulb Icon Allows user to turn on interior chamber light 19 Plug Icon Allows user to utilize interior electrical outlet 20 Temperature Display 3 digit LED indicates chamber temper...

Страница 16: ...und promptly report it to the carrier and request a damage inspection properly Important Failure to request an inspection of damage within a few days after receipt of shipment absolves the carrier fro...

Страница 17: ...Make sure the green ground wire is secured to the plug ground terminal Warning Do not operate shaker with a damaged electrical cord s To eliminate hazard of electrical shock make sure floor around sh...

Страница 18: ...g it to the wider bottom of the clamp The spring will hold the neck of the flask securely in place and provide security during shaking 2 Make sure all vessels are securely clamped before turning on un...

Страница 19: ...haker on 2 Press power switch a second time to turn shaker off 1 Hold down appropriate arrow key in the speed module of the control panel up or down until desired speed is set up to 500 rpm SET RPM li...

Страница 20: ...right side of the unit 2 Allow unit to stabilize at setpoint temperature before adjusting OTP 3 Once setpoint is stable rotate screwdriver counterclockwise until OSP appears in the Speed Display the T...

Страница 21: ...and system load it varies from minutes to hours 4 Once the chamber temperature has settled out record the temperature displayed by the independent sensor To obtain the best result an averaged temperat...

Страница 22: ...e shaker to reach speed 5 Using the SPEED UP and SPEED DOWN keys adjust the speed display to match the independent RPM sensor reading 6 Press the POWER key This completes shaking speed calibration 1 P...

Страница 23: ...cedures for configuring Microsoft Hyperterminal running on a host computer using Windows XP These instructions may need to be modified to be used with a different terminal emulator program and or oper...

Страница 24: ...systems 1 Rotate low limit control clockwise 2 Set chamber temperature using TEMP UP and DOWN arrow keys 3 Allow sufficient time for chamber temperature to stabilize before setting the low limit contr...

Страница 25: ...adjustment depending on environmental conditions to eliminate frost build up on the coils Refrigerated MaxQ 6000 units incorporate a defrost cycle to alleviate issues associated with the build up of c...

Страница 26: ...gh the evaporator coil to defrost it 9 Using the TEMP UP and TEMP DOWN arrow keys adjust the defrost duration to the desired time 10 Press the POWER key This logs the defrost duration and ends defrost...

Страница 27: ...perature display reading either rEF or Inc depending upon model state 3 Using the TEMP UP and TEMP DOWN keys adjust the model type to either rEF or Inc as desired 4 Press the POWER key This logs the m...

Страница 28: ...till advised especially when the repair may be technically sophisticated or difficult If assistance is needed beyond what the distributor can provide call Technical Services No merchandise should be r...

Страница 29: ...disconnect power cord 2 Remove top 3 screws from control panel nose 3 Remove air filter and clean Reinstall air filter 4 Replace control panel nose onto unit Make sure the studs located on the bottom...

Страница 30: ...s any key on control panel Check if continual timer is flashing Reset timer Platform doesn t rotate or has erratic speed Check for power to motor Replace motor if defective Check drive belt Replace if...

Страница 31: ...ower to the element removed OTP RTD temperature is above the OTP potentiometer setting Increase OTP potentiometer set ting OTP RTD temperature is above the OTP potentiometer setting Reduce Chamber set...

Страница 32: ...nt Shaking platform load too large Obstruction between shaking platform and chamber wall Remove obstruction bAL Shaker balance error The speed display will show bAL Unit vibrates and pro duces excessi...

Страница 33: ...present the unit will automatically clear them Errors E12 E13 and bAL are shaking errors Error Handler will lock out shaking functions if shaking error is detected Pressing any button or unplugging t...

Страница 34: ...MaxQ 6000 Shaker 8 1 Thermo Scientific...

Страница 35: ...8 2 MaxQ 6000 Shaker Thermo Scientific...

Страница 36: ...Motor 370 390 00 370 390 00 370 390 00 370 390 00 Motor Chamber Blower FA1415X1 FA1415X2 FA1415X1 FA1415X2 Motor Condenser MT1415X1 MT1415X2 PC Board Control PC1415X1 PC1415X1 PC1415X1 PC1415X1 PC Bo...

Страница 37: ...288 00 150 288 00 Circuit Breaker 1A 330 158 00 330 158 00 330 158 00 330 158 00 Compressor PUX36 PUX35 Element EL1415X1 EL1415X2 EL1415X1 EL1415X2 Fuse 230234 261575 230234 261575 Fuse FZX49 FZX49 GF...

Страница 38: ...warranty Extended warranties are dependent on the units being maintained regularly as stated in the operation and service manuals Replacement or repair of components parts or equipment under this warr...

Страница 39: ...i e glass filters light bulbs and lid gaskets are excluded from this warranty Extended warranties are dependent on the units being maintained regularly as stated in the operation and service manuals...

Страница 40: ...mo Fisher Scientific and its subsidiaries Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for deta...