6
Control Inspection
QC RUN
6-2
Sorvall RC 3BP Plus
Thermo Scientific
QC RUN
The QC RUN feature allows simple, automatic quality control run speed/temperature verification The
QC RUN feature will run and document the data from each set of parameters that has been saved in
program memory. In a QC RUN the centrifuge will start running program number 1, and
systematically run each program in order up to program number 9.
To minimize the time it takes to perform a QC RUN, we recommend the following:
•
Save the programs to the program memory so that the program 1 has the highest temperature. The
program temperatures should be lower the higher the program number is. If several programs use
the same temperatures, sort them by speed. The lowest speed should come first.
•
All programs must use the same rotor.
•
No program should use HOLD.
EXAMPLE:
The programs below are listed in correct order for using QC RUN (program values shown
are for example purposes only, actual set parameters will vary).
•
While the QC RUN is in progress, the parameters of the program that is running will be in the
SET DISPLAY.
•
During a QC RUN the centrifuge checks if the preset speed and temperature can be reached. In
order to save time during the QC RUN the next program will be run once the desired speed has
been reached and the temperature is within
° C of the preset value.
•
To select the QC RUN feature:
1. Press the
key.
2. Press the
key.
3. Install the correct rotor.
4. Press the
key.
The QC RUN sequence will begin, running program number 1.
ROTOR
SPEED
TIMED
DEG° C
PROG #
H-6000
2900
04:00 AM
22
1
H-6000
4000
03:00 AM
22
2
H-6000
3800
03:45 AM
4
3
H-6000
4700
05:30 AM
4
4
Note
If, during a QC RUN, the observed RUN temperature is more than 2° C over SET
temperature, the rotor continues spinning until set temperature is reached.
2