DB-066 E
2-7
Iw/Tr/Ff 06.07.2015
The alarm thresholds for dose rate and count rate are around
200 - 500 % above the typical normal background and allow
sensitive, but false alarm free dose rate monitoring with a rea-
sonable fast response time. The alarm thresholds for the dose
are set to the maximum possible values, thus being deactivat-
ed. With the setting of the history recording interval of 120 s ,
the last 52 hours of operation will be stored in the history
memory.
These default values can be changed by the user with the menu
functions (chapter 3.1) or with the help of the optional PC-
program „RadEye.EXE“. Furthermore, additional monitoring
modes can be activated.
Complete tables of factory settings according to the available
entries in the configuration RadEye.EXE software are provid-
ed in chapter 12-3.
The calibration parameters can only be changed in the factory
using special software tools.