5
Thermo Scientific Orion VERSA STAR Conductivity/Temperature Module
Chapter 3 Setup Menus
Setup Menus for Primary Measurements
1. Select the Conductivity channel by pressing the channel key, then press
f3 (Setup)
.
2. Arrow over to highlight the appropriate conductivity module for setup, press
f3 (Select)
.
3. Arrow the cursor to Mode, press
f3 (Select)
, arrow to Conductivity, press
f3 (Select)
.
4. Select the options for your testing.
TDS, Salinity and Resistivity Setup Menus
Conductivity Setup Menu
(Default values are in bold in the table below)
Probe Serial Number
Input
Sample ID
Off
, Manual, Auto Increment
Stability
Smart
, Fast, Medium, Slow
Averaging
Off,
Automatic Smart
Read Type
AutoRead
, Timed, Continuous
Cell K (Constant)
Input cell constant
Reference Temperature
5, 10, 15, 20,or
25
ºC
Temp Compensation
Linear
, nLFn, nLFu, EP
Temperature Coefficient
2 .1
Alarm**
Limit, CalDue, Set Point
** Limit and set point alarm default set to ‘off’.
Cal due alarm set to 12 hours.
TDS Setup Menu (Default values in bold)
Read Type
AutoRead
, Timed, Continuous
Cell K (Constant)
Input cell constant
TDS Factor
0 .50
Reference Temperature
5, 10, 15, 20, or
25
ºC
Temp Compensation
Linear
, nLFn, nLFu, EP, Off
Probe Serial Number
Input
Stability
Smart
, Fast, Medium, Slow
Averaging
Off,
Automatic Smart
Alarm
Limit, Set Point
Cell Type
Standard
, USP
Resistivity Setup Menu (Default values in bold)
Read Type
AutoRead, Timed, Continuous
Cell K (Constant)
Input cell constant
Reference Temperature
5, 10, 15, 20, or 25 ºC
Temp Compensation
Linear, nLFn, nLFu, EP, Off
Stability
Smart
, Fast, Medium, Slow
Averaging
Off,
Automatic Smart
Alarm
Limit, Set Point
Cell Type
Standard
, USP
Salinity Setup Menu (Default values in bold)
Read Type
AutoRead
, Timed, Continuous
Cell K (Constant)
Input cell constant
Salinity Type
Practical Salinity
, Sea Water
Stability
Smart
, Fast, Medium, Slow
Averaging
Off,
Automatic Smart
Alarm
Limit, Set Point
Cell Type
Standard
, USP
Содержание Orion Versa Star
Страница 1: ...Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Reference Guide...
Страница 4: ...This page intentionally left blank...
Страница 6: ...This page intentionally left blank...
Страница 12: ...8 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 14: ...10 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 16: ...12 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 18: ...14 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 20: ...16 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 34: ...30 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Notes...
Страница 35: ...This page intentionally left blank...