Thermo Scientific Orion VERSA STAR Conductivity/Temperature Module

Read Type Information

There are three main read types available as shown in the setup menus.

• 

Auto-Read – 

The meter will display the measurement as it stabilizes and lock and hold the 

measurement when it is stable. (The AR icon will lock on the screen.) Once the measurement is locked, 
the meter will automatically export the measurement to the data log, if the data log is enabled in 
the setup menu, and to a printer or computer, if a printer or computer is connected to the meter and 
enabled in the setup menu. To take a new measurement, press 

measure (esc)/1

.

• 

Timed – 

The meter will read and display a measurement at the set time interval that is programmed 

by the operator. There are two options: 

 

1. 

By checking/selecting the "Interval" box, at the set time interval, the meter will automatically  

 

export the measurement to the data log, if the data log is enabled in the setup menu, and to a 

 

 

printer or computer, if a printer or computer is connected to the meter and enabled in the setup 

 

 

menu. If selected, enter the time interval value in a hours:minutes:seconds format using the 

 

 

numeric keypad and press 

f3 (select)

 key.

 

2. 

 If the "Interval" box is unchecked/not selected, a single timed measurement will be taken.  

 

The meter will apply the same saving and exporting data conditions as with the set time 

 

 

interval to record and export data, but lock after one timed interval has expired. Press 

 

 

measure (esc)/1

 to reset the countdown clock and take a new reading.

• 

Continuous –

 The meter will continuously measure and update the display. This read type is useful 

when performing an experiment that requires continuous measurements to be taken. Press the 

 

log/print/0

 key to export the measurement to the data log, if the data log is enabled in the setup 

menu, and to a printer or computer, if a printer or computer is connected to the meter and enabled in 

the setup.

Temperature Compensation Options

These options can vary according to the measurement mode.

• 

Linear –

 Most beneficial for use with samples with known and linear temperature coefficients.

• 

nLFn –

 This abbreviation is for non-linear natural pure water. It works best for low conductivity water 

such as pure waters and natural waters (ground, well or surface waters).

• 

nLFu –

 This abbreviation is for non-linear ultra pure water that contains no air or carbon dioxide.

• 

EP –

 Temperature-compensation is turned off when this is selected. Using this feature results in 

warning limits being displayed when the pure water sample is over the acceptable reading for that 
temperature per EP or USP requirements.

Содержание Orion Versa Star

Страница 1: ...Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Reference Guide...

Страница 2: ...riode Minimum Stir Requirement MSR NISS Optimum Results Orion Dual Star Orion Star SAOB SMART AVERAGING SMART STABILITY Star LogR Star Navigator 21 Stat Face Triode are trademarks of Thermo Fisher Sci...

Страница 3: ...rement 13 Chapter 7 Methods 15 Chapter 8 Data Storage and Review 17 Data Storage Settings 17 Measurement Read Type 17 Datalog Computer and Printer Export Settings 17 Datalog Review 17 Calibration Revi...

Страница 4: ...This page intentionally left blank...

Страница 5: ...f this module include Conductivity TDS salinity and resistivity capabilities Selectable reading reference temperatures of 5 10 15 20 or 25 C Compatibility with 2 or 4 cell conductivity cells Linear no...

Страница 6: ...This page intentionally left blank...

Страница 7: ...sample range and reporting requirements 2 Place conductivity cell in the meter s electrode arm 3 Attach the 8 pin mini DIN connection into the conductivity module 4 Attach the Orion Star stirrer prob...

Страница 8: ...tance using the proper cleaning procedure for that spill type Display Information from Module Example Conductivity Measurement Screen Single Channel Date and time Sample temperature Primary reading In...

Страница 9: ...2 1 Alarm Limit CalDue Set Point Limit and set point alarm default set to off Cal due alarm set to 12 hours TDS Setup Menu Default values in bold Read Type AutoRead Timed Continuous Cell K Constant I...

Страница 10: ...ox is unchecked not selected a single timed measurement will be taken The meter will apply the same saving and exporting data conditions as with the set time interval to record and export data but loc...

Страница 11: ...mperature probe to be used in more than one channel for temperature compensated measurements such as when a pH electrode and a 2 cell conductivity cell are in the same measurement sample Temperature i...

Страница 12: ...8 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Страница 13: ...ll user guide Note For automatic calibration the nominal cell constant of the conductivity cell must be entered in the setup menu before the calibration is performed Use fresh Thermo Scientific Orion...

Страница 14: ...10 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Страница 15: ...ents recalibrate if a different conductivity cell is used For temperature calibration the probe needs to be connected to the meter and the calibration solution should have a known stable temperature I...

Страница 16: ...12 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Страница 17: ...t sample 3 If the meter is in AutoRead mode meter default press measure esc 1 key If the meter is in continuous or timed read mode the meter will immediately start taking readings 4 If in AutoRead mod...

Страница 18: ...14 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Страница 19: ...he last calibration data stored for each method Use the channel setup to access and modify methods 1 Press f3 setup key 2 Use the arrow keys to highlight the appropriate channel Press f3 select 3 Use...

Страница 20: ...16 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...

Страница 21: ...ke your selection 1 From measurement mode press setup 3 2 Press the arrow keys to highlight Inst Settings Press setup 3 3 Use the arrow keys f2 page key and f3 select key to a set the communication pr...

Страница 22: ...of the last 10 calibrations for each parameter of the meter The information will match the information recorded for that calibration To review calibration information 1 Press log view 9 key 2 Press ri...

Страница 23: ...s in your country and information on Thermo Scientific Orion products that may assist the detection of substances subject to the RoHS Directive are available at www thermoscientific com Error Recommen...

Страница 24: ...A Equipment Class Measurement control and laboratory Orion Star A series meters are EMC Class A Conforms with the following directives and standards EN61326 1 2006 Electromagnetic Compatibility EMC Di...

Страница 25: ...ation Linear 0 to 10 0 C nLn nLFu EP Compatible Cell Constants 0 001 to 199 9 Calibration Points Up to 6 points Calibration Editing Yes Resistivity Range 2 ohm to 100 meg ohm Resolution 2 ohms cm Rela...

Страница 26: ...Set Orion VERSA STAR meter Orion VERSA STAR conductivity temperature module Electrode arm with redesigned holder Universal power adapter VSTAR22 Orion VERSA STAR Conductivity Benchtop Meter Kit Orion...

Страница 27: ...th Two pH ISE Modules Kit Orion VERSA STAR meter Two Orion VERSA STAR pH ISE temperature modules 8102BNUWP Orion ROSS Ultra pH electrode 096019 Orion Star stirrer probe 927007MD ATC probe stainless st...

Страница 28: ...bottle 810007 ROSS fill solution 5x60 mL Electrode arm with redesigned holder Universal power adapter VSTAR90 Orion VERSA STAR pH ISE Conductivity RDO Dissolved Oxygen Benchtop Multiparameter Meter S...

Страница 29: ...rage bottle 011007 conductivity standard 1413 S 5x60 mL Electrode arm with redesigned holder Universal power adapter VSTAR93 Orion VERSA STAR pH ISE Conductivity RDO Dissolved Oxygen Ultimate Benchtop...

Страница 30: ...body 927005MD Orion ATC probe epoxy body 013005MD Orion DuraProbe conductivity cell 4 cell K 0 475 013016MD Orion conductivity cell flow through K 0 1 083005MD Orion dissolved oxygen probe polarograp...

Страница 31: ...f many aqueous samples Solution 25 C to 50 C Temperature Coefficient C Ultra Pure Water 4 55 Salt NaCl 2 12 5 NaOH 1 72 Dilute Ammonia 1 88 10 HCl 1 32 5 Sulfuric Acid 0 96 98 Sulfuric Acid 2 84 Sugar...

Страница 32: ...99 83 56 2 68 59 7 555 822 86 58 3 70 35 7 768 846 88 59 4 72 12 7 983 870 91 61 5 73 91 8 200 894 93 63 6 75 70 8 418 918 96 64 7 77 50 8 638 943 98 66 8 79 32 8 860 968 101 68 9 81 15 9 084 992 103...

Страница 33: ...85 1636 170 117 34 130 4 15 10 1665 173 119 35 132 5 15 35 1693 176 121 36 134 6 15 61 1722 179 123 37 136 7 15 86 1751 182 125 38 138 9 16 12 1780 185 127 39 141 0 16 37 1808 188 129 40 143 2 16 63 1...

Страница 34: ...30 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Notes...

Страница 35: ...This page intentionally left blank...

Страница 36: ...thermo com India Tel 91 22 4157 8800 wai asia thermofisher com Japan Tel 81 045 453 9175 wai asia thermofisher com China Tel 86 21 68654588 wai asia thermofisher com Singapore Tel 65 6778 6876 wai asi...

Отзывы: