![Thermo Scientific Orion Versa Star Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/thermo-scientific/orion-versa-star/orion-versa-star_reference-manual_1104611017.webp)
13
Thermo Scientific Orion VERSA STAR Conductivity/Temperature Module
Chapter 6 Measurement
After successful calibration, the samples can be tested.
1. Rinse the conductivity cell with deionized water, blot dry with lint-free tissue.
2. Place conductivity cell and stirrer in first sample.
3. If the meter is in AutoRead mode (meter default) press
measure (esc) (1)
key. If the meter is in
continuous or timed read mode, the meter will immediately start taking readings.
4. If in AutoRead mode and datalog is enabled (default), the reading will automatically be stored when
the “AR” appears. If in continuous read mode and datalog is enabled, press
log/print (0)
key to store
into the meter’s memory.
Note:
To enable or disable datalogging, press
setup
and select Inst. Settings.
Содержание Orion Versa Star
Страница 1: ...Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Reference Guide...
Страница 4: ...This page intentionally left blank...
Страница 6: ...This page intentionally left blank...
Страница 12: ...8 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 14: ...10 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 16: ...12 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 18: ...14 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 20: ...16 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module This page intentionally left blank...
Страница 34: ...30 Thermo Scientific Orion VERSA STAR Conductivity Temperature Module Notes...
Страница 35: ...This page intentionally left blank...