24
Thermo Scientific Orion Star A329 Portable pH/ISE/Conductivity/RDO/DO Meter Reference Guide
Calibration Editing
*OUIFDBMJCSBUJPOTVNNBSZEJTQMBZ *4&DBMJCSBUJPOTUFQQSFTT
f3 (cal edit)
to edit the calibration points.
2. Press
or
to highlight the calibration point to be edited and press
f2 (select)
.
3. Press
or
to highlight
Remeasure
,
Edit
or
Delete
and press
f2 (select).
a. If
Remeasure
JTTFMFDUFESFQFBU*4&DBMJCSBUJPOTUFQTUISPVHI
b. If
Edit
is selected, press
f3 (edit)
UPBDDFTTUIFOVNFSJDFOUSZTDSFFOFOUFSUIFOFXWBMVFQSFTT
f2 (done)
and then press
f2 (accept)
. Select another calibration point to edit or press
f1 (back)
.
c. If
Delete
JTTFMFDUFEUIFDBMJCSBUJPOQPJOUXJMMCFEFMFUFE*GEFMFUJOHUIFDBMJCSBUJPOQPJOUDIBOHFT
the calibration to a one point calibration, enter the slope using the numeric entry screen.
5IFNFUFSXJMMEJTQMBZBOVQEBUFEDBMJCSBUJPOTVNNBSZBOEFYQPSUUIFEBUBUPUIFDBMJCSBUJPOMPH1SFTT
f1 (meas)
to proceed to the measurement mode, press
f2 (print)
to proceed to the measurement mode and
export the data to a printer or computer or press
f3 (cal edit)
to edit another calibration point.