Thermo Scientific Orion Star A329 Portable pH/ISE/Conductivity/RDO/DO Meter Reference Guide
Chapter 5 ISE Calibration
ISE Calibration
0OFUPmWFTUBOEBSETDBOCFVTFEGPS*4&DBMJCSBUJPO*GNPSFUIBOPOFTUBOEBSEJTVTFEUPDBMJCSBUJPOTUBSUXJUIUIF
MPXFTUDPODFOUSBUJPOTUBOEBSEBOEXPSLVQUPUIFIJHIFTUDPODFOUSBUJPOTUBOEBSEMBTU"MXBZTVTFGSFTITUBOEBSET
Select standards that bracket the sample concentration and are a decade apart in concentration. Prepare the ion
selective electrode according to the instructions in the electrode use guide. Connect the ISE and any other electrodes
UPCFVTFE "5$QSPCFSFGFSFODFFMFDUSPEFUPUIFNFUFS1PXFSPOUIFNFUFSBOETFUUIFNFBTVSFNFOUNPEFUP*4&
Note: In most calibration screens, press
f1 (esc)
to return to the measurement mode without saving the calibration.
*OUIFNFBTVSFNFOUNPEFQSFTT
f1 (cal).
In dual and multi channel measurement displays, press
or
to highlight
pH/ISE – Channel
and press
f2 (select)
.
3JOTFUIFJPOTFMFDUJWFFMFDUSPEFBOEBOZPUIFSFMFDUSPEFTJOVTFXJUIEJTUJMMFEXBUFSCMPUESZXJUIB
MJOUGSFFUJTTVFBOEQMBDFJOUPUIF*4&TUBOEBSE
3. When the electrode and standard are ready, press
f3 (start)
.
8BJUGPSUIFDPODFOUSBUJPOWBMVFPOUIFNFUFSUPTUBCJMJ[FBOETUPQnBTIJOHBOEQFSGPSNPOFPGUIFGPMMPXJOHBDUJPOT
a. Press
f2 (accept)
to accept the displayed value.
b. Press
f3 (edit)
to access the numeric entry screen and edit the value.
i. Press
,
,
or
to highlight a number or decimal point; press
f3 (enter)
to select
UIFIJHIMJHIUFEJUFNBOESFQFBUVOUJMUIFWBMVFBUUIFNFBTVSFEUFNQFSBUVSFJTTIPXOBCPWFUIF
numeric entry screen.
ii. Press
f2 (done)
to exit the numeric entry screen.
iii. Press
f2 (accept)
to accept the entered value.
1SFTT
f2 (next)
UPQSPDFFEUPUIFOFYUTUBOEBSEBOESFQFBUTUFQTUISPVHIPSQSFTT
f3 (cal done)
to save and
FOEUIFDBMJCSBUJPO*GmWFTUBOEBSETBSFVTFEUIFDBMJCSBUJPOXJMMTBWFBOEFOEPODFUIFmGUIWBMVFJTBDDFQUFE
a. If a one point calibration is performed, press
f2 (accept)
to accept the displayed slope value or
press
f3 (edit)
to access the numeric entry screen, enter the slope value and press
f2 (accept)
.
5IFNFUFSXJMMEJTQMBZUIFDBMJCSBUJPOTVNNBSZJODMVEJOHUIFBWFSBHFTMPQFBOEFYQPSUUIFEBUBUPUIFDBMJCSBUJPO
log. Press
f1 (meas)
to proceed to the measurement mode or press
f2 (print)
to proceed to the measurement
mode and export the data to a printer or computer.