5
Connecting the Cables and Power Cords
16
Orbitrap Tribrid Series Getting Connected Guide
Thermo Scientific
To connect the MS and data system cables and power cords
1
1
The back of the data system computer might not be as shown.
(Not shown) Connect both ends of the
power
cords
(for the computer, Ethernet switch, modular
valve, syringe pump, MS, and forepumps).
IMPORTANT
Plug the power cords for the MS
and forepumps into separate, dedicated 230 Vac
electrical outlets—that is, circuits A, B, and C for a
dual-forepump system.
Note
If the modular valve’s LED remains off, press
its front Remote button to exit local mode.
5
1
Connect the
relay
control cable
between
the MS (V65 Pump
Enable) and the
forepumps.
4
(Not shown) Connect the
Ready Out
cable
(not provided) and
contact
closure cable
and Ground
pins) between the MS and the external
device. Then, attach the
connector
housings
. See
2
Connect the
USB cables
from the
MS to the syringe pump and the
modular valve.
3
Connect the
Ethernet cables
from the
Ethernet switch to the computer’s LC/MS
Ethernet network card and the MS.
IMPORTANT
Before turning on the MS, make sure to complete all of the system
connections described in this guide.
Tip
For troubleshooting purposes, you might want to record which devices connect to
which ports on the Ethernet switch.