The Filter Config Screen
1-84
NITON XL3 Analyzer User’s Guide
Thermo Scientific
The Filter Config
Screen
Figure 1-11. The Filter Config Menu Path
Multi-Filter tests are used to either preferentially excite specific elements for
increased sensitivity, or to cover a wider element range than one filter alone
can provide. Most modes, when enabled, will use two filters in sequence to
produce a combined analysis result. In typical alloy analysis applications,
Main Range is used for the analysis of most elements, and Low Range is
utilized for the subsequent high sensitivity analysis of V, Ti, and Cr.
Multi-filter switching can be set to activate off time alone, or, when time
switching is disabled, off settings in the alloy grade library. In environmental
modes, Low Range adds the capability to analyze light elements which
cannot be efficiently excited by Main Range.
Содержание NITON XL3t 500
Страница 1: ...NITON XL3t 500 Analyzer User s Guide Version 6 5 ...
Страница 2: ......
Страница 22: ...Applications Live Spectrum Feed 1 8 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 40: ...The Measurement Screen 1 26 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 66: ...The View Data Screen 1 52 NITON XL3 Analyzer User s Guide Thermo Scientific Figure 1 8 The SpectraView Screen ...
Страница 120: ...The Sort Element Display Menu 1 106 NITON XL3 Analyzer User s Guide Thermo Scientific Figure 1 40 Save Changes ...
Страница 160: ...Entering Data with a Barcode reader 2 32 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 182: ...Radiation and General Safety Contact Information 3 22 NITON XL3 Analyzer User s Guide Thermo Scientific ...
Страница 200: ...Appendix F Appendices xviii NITON XL3 Analyzer User s Guide Thermo Scientific ...