![Thermo Scientific Nicolet Summit OA Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/thermo-scientific/nicolet-summit-oa/nicolet-summit-oa_user-manual_1105202050.webp)
Calibrate the Laser Frequency
Calibrate your laser frequency if the Nicolet Summit OA - PV Test workflow indicates a failing test.
When a test fails, first align the spectrometer, then calibrate the laser frequency and run the test
again. If the test still fails, contact your local technical support representative for help.
❖
To calibrate the laser (touchscreen interface)
1. On the home screen, select the Diagnostics icon [
] to open the Diagnostics view.
2. Open the Laser Calibration tab and select Calibrate to begin.
When laser calibration is complete, a message indicates whether the laser frequency has been
changed and displays the new frequency.
❖
To calibrate the laser (operator interface)
1. Navigate to the Administrator area, and select the Diagnostics icon [
] to open the Dia-
gnostics view.
2. Open the Laser Calibration tab and select Calibrate to begin.
When laser calibration is complete, a message indicates whether the laser frequency has been
changed and displays the new frequency.
❖
To calibrate the laser (desktop interface)
1. Select Acquire Data > Diagnostics > Laser Calibration.
2. In the Laser Calibration dialog box, click Start to calibrate the laser.
When the system is finished calibrating the laser, a message indicates whether the frequency
was changed and displays the new frequency.
Thermo Scientific
Nicolet Summit OA Spectrometer User Guide
47
Maintenance
Содержание Nicolet Summit OA
Страница 7: ... This page intentionally left blank ...
Страница 45: ... This page intentionally left blank ...