![Thermo Scientific Nicolet Summit OA Скачать руководство пользователя страница 49](http://html.mh-extra.com/html/thermo-scientific/nicolet-summit-oa/nicolet-summit-oa_user-manual_1105202049.webp)
Maintenance
Align the Spectrometer Optics
Align your spectrometer’s optics if the Nicolet Summit OA - PV Test workflow indicates a failing test.
Aligning the spectrometer optimizes the energy that reaches the detector and maximizes the detector
signal. If the performance test fails, align the spectrometer’s optics, calibrate the laser frequency, and
run the test again. If the test still fails, contact your local technical support representative for help.
❖
To align the spectrometer optics (touchscreen interface)
1. On the home screen, select the Diagnostics icon [
] to open the diagnostics view.
2. Open the Align tab and select Align.
The system automatically aligns the spectrometer and indicates when alignment is complete.
❖
To align the spectrometer optics (operator interface)
1. Navigate to the Administrator area, select the Diagnostics icon [
] to open the diagnostics
view.
2. Open the Align tab and select Align.
The system automatically aligns the spectrometer and indicates when alignment is complete.
❖
To align the spectrometer optics (desktop interface)
1. Select Acquire Data > Diagnostics > Align Spectrometer.
2. In the Align Spectrometer dialog, click Start.
The system automatically aligns the spectrometer and indicates when alignment is complete.
46
Nicolet Summit OA Spectrometer User Guide
Thermo Scientific
Содержание Nicolet Summit OA
Страница 7: ... This page intentionally left blank ...
Страница 45: ... This page intentionally left blank ...