Maintenance
Decontaminating the Instrument
296
NanoDrop One User Guide
Thermo Scientific
Related Topics
• PR-1 Pedestal Reconditioning Kit
Decontaminating the Instrument
Decontaminate the instrument after measurements with samples that contain
and before returning the instrument to us for maintenance or repair.
Tip
The PR-1 pedestal reconditioning compound is the easiest way to recondition the
pedestals. If you don’t have a PR-1 kit, follow these steps:
1. Lift the instrument arm and pipette 3 μL 0.5M HCl onto the lower pedestal.
2. Lower the arm and wait 2–3 minutes.
3. Lift the arm and clean both pedestals with a new laboratory wipe.
4. Pipette 3 μL DI H
2
O onto the lower pedestal.
5. Lower the arm and wait 2–3 minutes.
6. Lift the arm and clean both pedestals with a new wipe.
NOTICE
: Support the instrument arm with one hand while you buff the upper
pedestal to avoid damaging the arm.
7. Fold a clean laboratory wipe into quarters and use it to vigorously buff the surface of
each pedestal at least 50 times.
8. Use canned air to remove any paper residue from the pedestals.
Note
If your instrument requires maintenance or repair,
or your local
distributor.
Содержание NanoDrop One
Страница 1: ...NanoDrop Micro UV Vis Spectrophotometers NanoDrop One User Guide 269 309102 Revision A May 2017 ...
Страница 18: ......
Страница 28: ......
Страница 58: ......
Страница 134: ...13 Measure Protein Pierce 660 128 NanoDrop One User Guide Thermo Scientific Instrument Settings ...
Страница 144: ......
Страница 152: ......
Страница 159: ...16 Measure UV Vis Thermo Scientific NanoDrop One User Guide 153 ...
Страница 160: ...16 Measure UV Vis 154 NanoDrop One User Guide Thermo Scientific ...
Страница 193: ...18 Learning Center Set Up the Instrument Thermo Scientific NanoDrop One User Guide 187 Instrument Settings ...
Страница 293: ...18 Learning Center Multimedia Thermo Scientific NanoDrop One User Guide 287 Pedestal Cleaning and Reconditioning ...
Страница 294: ...18 Learning Center Multimedia 288 NanoDrop One User Guide Thermo Scientific ...
Страница 314: ...19 Maintenance Instrument Diagnostics 308 NanoDrop One User Guide Thermo Scientific ...
Страница 328: ...7 Technical Support 314 NanoDrop One User Guide Thermo Scientific ...