Section 17- Appendices
17. Appendices
Instrument Specifications
•
Sample Size: 1microliter
•
Path Length: 1 mm (with auto-ranging to 0.2 mm)
•
Light Source: Xenon flash lamp
•
Detector Type: 2048-element linear silicon CCD array
•
Wavelength Range: 220-750 nm
•
Wavelength Accuracy: 1 nm
•
Wavelength Resolution: 3 nm (FWHM at Hg 546 nm)
•
Absorbance Precision: 0.003 absorbance ( 1mm path)
•
Absorbance Accuracy: 2% (at 0.76 absorbance at 257 nm)
•
Absorbance Range: 0.02-75 (10 mm equivalent absorbance)
•
Detection Limit: 2 ng/microliter (dsDNA)
•
Maximum Concentration: 3700 ng/microliter (dsDNA)
•
Measurement Cycle Time: 10 seconds
•
Dimensions: 20 cm X 15 cm x 12 cm
•
Weight: 3 kg
•
Sample Pedestal Material of Construction: 303 stainless steel and quartz fiber
•
Operating Voltage: 12 Vdc
•
Operating Power Consumption: 6 W
•
Standby Power Consumption: 1.5 W
•
CE Approval: Units sold in Europe, Australia and New Zealand
•
UL/CSA Approval: Units sold in N. America, S. America, Asia and Africa
•
Included in system: software, compatible with Windows 2000 or XP
Blanking and Absorbance Calculations
When the NanoDrop
®
ND-1000 Spectrophotometer is “blanked”, a spectrum is taken of a reference material (blank) and stored in
memory as an array of light intensities by wavelength. When a measurement of a sample is taken, the intensity of light that has
transmitted through the sample is recorded. The sample intensities along with the blank intensities are used to calculate the sample
absorbance according to the following equation:
Absorbance = -log (Intensity
sample
/Intensity
blank
)
Thus, the measured light intensity of both the sample and of the blank are required to calculate the absorbance at a given wavelength.
Concentration Calculation (Beer’s Law)
General
The Beer-Lambert equation is used to correlate the calculated absorbance with concentration:
A =
E
*
b
*
c
Where
A
is the absorbance represented in absorbance units (A),
E
is the wavelength-dependent molar absorptivity coefficient (or
extinction coefficient) with units of liter/mol-cm,
b
is the path length in cm, and
c
is the analyte concentration in moles/liter or molarity
(M).
Fluorescent Dyes (Microarray Measurement)
The NanoDrop software uses the general form of the Beer-Lambert equation to calculate fluorescent dye concentrations in the
Microarray module. The table of extinction coefficients for each dye is below:
Dye Type
Extinction
Coefficient
(liter/mol-cm)
Measurement
Wavelength
(nm)
Cy3 150000
550
Cy5 250000
650
Alexa Fluor 488
71000
495
Alexa Fluor 546
104000
556
Alexa Fluor 555
150000
555
Alexa Fluor 594
73000
590
Alexa Fluor 647
239000
650
Alexa Fluor 660
132000
663
Cy3.5 150000
581
Cy5.5 250000
675
Nucleic Acids
17-1