Note
While adjusting the platform retaining screws, make certain lid
switch, located at back of chamber, is activated (pressed in).
3. While it is slowly orbiting, slightly tighten each of the A screws in
Figure 1. Repeat this procedure by slightly tightening all the B screws,
then complete procedure by tightening the C screws.
Caution
All screws must first be slightly tightened to seat plate properly.
s
4. Repeat the screw-tightening procedure at 50-100 RPM until all 9
screws are securely tightened.
Suggested with every 3 months of constant use.
Caution
Any internal adjustments or repairs must be performed by a
qualified service representative.
s
Remove the platform by loosening 4, thumbscrews in the platform center.
Remove the sheet metal panel (9 screws) under the platform to expose the
belt and interior parts. Inspect the drive belt for wear. Order a replacement
if necessary.
6-2
Incubated and Refrigerated Shakers
Thermo Scientific
Section 6
Maintenance
)&!04&1 %), 1"!1
(,# %*"/
"/&+$*,2+10
-)1#,/**,2+1&+$
-)1"
%
&
$
&
%
%
&
$
$
Figure 6-1.
Platform Mounting Plate
Platform
Maintenance
Replacing Platform
Mounting Plate
(cont.)